2. INSTALLATION
Page 11
© 2007 DH Instruments, a Fluke Company
1. 60 in. x 1/8 in. nipple with DH500 tips
(to GB-152-AF high pressure
OUT
port)
2. DH500 union
3. 6 in. DH500 nipple
4. DH500 elbow
5. 2.75 in. DH500 nipple
6. GPC1-10000-AF
SUPPLY
port
7. GPC1-10000-AF
Figure 5.
Connection of GB-152-AF high pressure OUT port to
GPC1-10000-AF SUPPLY port
Depending on the drive pressure set by the DRIVE ADJUST regulator,
GB-152-AF may generate pressure over 70 MPa (10 000 psi). To limit the maximum
pressure that can be generated by the booster, limit the drive air supply to the Drive
Air Control Kit to 75 psi (500 kPa) (see Section 2.3.2.1). Be sure to connect the
booster high pressure output using fittings and tubing rated for the maximum
expected pressure that is to be output from the booster. Always use a DH500 or
equivalent tube on the DH500 F high pressure OUT port connection.