ASCII Programming
Command List
127
14
Description
Char
Value Format
Poll Burst Set Legal Values
Factory Default
Error Codes
EC
nnnnnnnnnnnnnnnn
√
√
2
15
alarm detection
2
14
narrow band temp. over range
2
13
narrow band temp. under
range
2
12
wide band temp. over range
2
11
wide band temp. under range
2
10
two-color temp. over range
2
9
two-color temp. under range
2
8
attenuation>95% (dirty window)
2
7
attenuation too high (> 95%)
2
6
energy too low
2
5
narrow band detector failure
2
4
wide band detector failure
2
3
internal temp. under range
2
2
internal temp. over range
2
1
heater control temp. under
range
2
0
heater control temp. over range
Emissivity source
ES
X
√
I = set by a constant number
according to the command "E"
E = set by the analog input
I
Valley hold time (1)
F
n.n - nnn.n
√
√
√
0.0
– 300.0 s (300 s =
)
000.0
Average time (1)
G
n.n - nnn.n
√
√
√
0.0
– 300.0 s (300 s =
)
000.0
Gateway Address
GW
nnn.nnn.nnn.nnn
√
√
0.0.0.0 - 255.255.255.255
192.168.42.1
Top of mA temperature range
H
nnnn.n
– nnnn.n
√
√
√
min/max of temperature range (°C
or °F)
High
temperature
range of sensor
RS485 mode
HM
n
√
√
2 = 2 wire
4 = 4 wire
2
Sensor internal ambient
I
n.n - nnn.n
√
√
0°C/32°F
– 65°C/149°F
Analog input mA, current value IN
nn.nn
– nn.nn
√
√
0 - 20 or 4 - 20
Analog input mode
INM
n
√
√
0 = 0
– 20 mA
4 = 4
– 20 mA
0
IP Address
IP
nnn.nnn.nnn.nnn
√
√
0.0.0.1 - 255.255.255.255
192.168.42.132
Switch panel lock
J
X
√
√
L = Locked
U = Unlocked
Unlocked
Relay alarm output control
K
n
√
√
0 = Permanently Open
2
1 = Permanently Closed
2 = Normally Open
3 = Normally Closed
Bottom of mA temperature
range
L
n.n
– nnnn.n
√
√
√
0.0
– 9999.0 (°C or °F)
Low temperature
range of sensor
Color mode
M
n
√
√
√
1 = 1 - color
2
2 = 2 - color
MAC Hardware Address
MAC
nnnnnnnnnnnn
√
e.g. 001d8d2aaa01
Set at factory
calibration
Target temp:1C narrow
N
n.n - nnnn.n
√
√
Net Mask
NM
nnn.nnn.nnn.nnn
√
√
0.0.0.1 - 255.255.255.255
255.255.255.0
Output current
O
nn
√
√
√
00 = controlled by unit
00
02 = under range
21 = over range
0
– 20 = current in mA
Output format
OIF
0 = temperature value format is
float with one decimal place
1 = temperature value format is
integer with four digits
0
Analog output gain
OUG
n.nn
√
√
0.01 to 100.0
1.0
Analog output offset
OUO
-nnn - +nnn
√
√
-200 to +200
0.0
Peak hold time (1)
P
n.n - nnn.n
√
√
√
0.0
– 300.0 s (300 s =
)
0.0
ProfiNet Name
PNN
√
√
IP Port address
PORT
n - nnnnn
√
√
1 - 65535
6363
Wide Power
Q
n.nnnnnnn
√
√
Narrow power
R
n.nnnnnnn
√
√