DTX-PCU6S Patch Cord Adapters
Users Manual
A-2
To test the same adapters later, do the following:
1
Connect the reference patch cord between the same main and remote testers used
for the original test. Be sure the marked ends of the patch cord go to the correct
testers.
2
Run an Autotest; then save the results.
3
Compare the results with the reference patch cord’s previous results. The difference
between the old and new NEXT and RL results should be less than 0.5 dB near the
PASS/FAIL limits.
If the old and new results are more than 0.5 dB apart near the PASS/FAIL limits, the RJ45
jacks may need replacing. See “Replacing the Adapter's Jacks” on page 8 for details.