9000A-006 Service
3-2
•
Perform all repairs at a static-free work station.
•
Do not handle the contact points of the SMDs.
•
Attach grounding straps to repair personnel.
•
Remove all plastic, vinyl, and styrofoam from the work area.
DISASSEMBLY PROCEDURE
3-6.
The Asynchronous Signature Module is located within the Troubleshooter case. Gain
access to and remove the module as follows:
1. Invert the instrument on a clean surface, and remove the seven retaining screws
from the bottom side.
2. Carefully return the instrument to the upright position while holding the top cover
in place. Once the instrument is in the upright position, remove the top cover and
lay it to the right side of the instrument, exposing the Signature Module as shown
in Figure 3-1.
3. If it is necessary to remove the module from the base assembly of the Trouble-
shooter, proceed as follows:
a. Remove the three screws that secure the Signature Module PCB assembly
within the Troubleshooter.
b. Lift out the module and disconnect the four ribbon cables that extend from the
module to the main assembly of the Troubleshooter.
FIGURE 3-1. Location of Asynchronous Signature Module
Summary of Contents for 9000A-006
Page 6: ...9000A 006 Service iv ...
Page 8: ...9000A 006 Service vi ...
Page 14: ...9000A 006 Service 1 6 ...
Page 27: ...2 13 9000A 006 Service FIGURE 2 6 Waveform Section Block Diagram ...
Page 38: ...9000A 006 Service 3 8 ...
Page 42: ...9000A 006 Service 4 4 FIGURE 4 1 9000A 006 Final Assembly ...
Page 43: ...4 5 9000A 006 Service Figure 4 1 9000A 006 Final Assembly cont ...
Page 44: ...9000A 006 Service 4 6 Figure 4 1 9000A 006 Final Assembly cont ...
Page 46: ...9000A 006 Service 4 8 FIGURE 4 2 A1 Main PCA ...
Page 48: ...9000A 006 Service 4 10 FIGURE 4 3 A2 Clock Module PCA ...
Page 54: ...9000A 006 Service 5 2 ...
Page 55: ...9000A 006 Service 5 3 Figure 5 1 A1 Main PCA ...
Page 56: ...9000A 006 Service 5 4 Figure 5 1 A1 Main PCA ...
Page 57: ...9000A 006 Service 5 5 Figure 5 1 A1 Main PCA cont ...
Page 58: ...9000A 006 Service 5 6 Figure 5 1 A1 Main PCA cont ...
Page 59: ...9000A 006 Service 5 7 Figure 5 1 A1 Main PCA cont ...
Page 60: ...9000A 006 Service 5 8 Figure 5 1 A1 Main PCA cont ...
Page 63: ...9000A 006 Service 5 11 Figure 5 2 A2 Clock Module Assembly ...
Page 64: ...9000A 006 Service 5 12 ...