87 V/AN
Service Manual
32
Notes
Set the calibrator to Standby prior to changing the function switch position
and or after completing adjustment of each function.
If the calibration adjustment procedure is not completed correctly, the
Meter will not operate correctly.
Table 16. Calibration Adjustment Steps
Switch Position
(Function)
Input
Terminal
Calibration Adjustment
Step
Input Value
C-01
600.0 mV, 60 Hz
C-02
600.0 mV, 20 kHz
C-03
6.000 V, 60 Hz
C-04
6.000 V, 20 kHz
C-05
60.00 V, 60 Hz
C-06
60.00 V, 20 kHz
C-07
600.0 V, 60 Hz
K
(AC Volts)
C-08
600.0 V, 10 kHz
C-09
6.000 V, 0 Hz
C-10
60.00 V, 0 Hz
L
(DC Volts)
C-11
600.0 V, 0 Hz
C-12
600.0 mV, 0 Hz
m
L
(DC Millivolts)
C-13
60.00 mV, 0 Hz
C-14 600.0
Ω
C-15 6.000
k
Ω
C-16 60.00
k
Ω
C-17 600.0
k
Ω
C-18 6.000
M
Ω
C-19 0.000
Ω
Ω
(Ohms)
C-20 50.0
M
Ω
O
(Diode Test)
I
C-21
3.000 V, 0 Hz
C-22
6.000 A, 60 Hz
A
C-23
6.000 A , 0 Hz
C-24
60.00 mA, 60 Hz
C-25
400.0 mA, 60 Hz
C-26
60.00 mA, 0 Hz
A, mA
(Amps, milliamps)
mA /
U
A
C-27
400.0 mA, 0 Hz
C-28 600.0
µ
A, 60 Hz
C-29 6000
µ
A, 60 Hz
C-30 600.0
µ
A, 0 Hz
U
A
(Microamps)
mA /
U
A
C-31 6000
µ
A, 0 Hz