Required Equipment
Required equipment for the performance tests is listed in Table 13. If the recommended
models are not available, equipment with equivalent specifications may be used.
XW
Warning
•
To avoid shock or injury, do not perform the verification
tests or calibration adjustment procedures described in this
manual unless you are qualified to do so.
•
Repairs or servicing should be performed only by qualified
personnel.
Table 13. Required Equipment
Equipment
Required Characteristics
Recommended Model
Calibrator
AC Voltage Range: 0 - 1000 V ac
Accuracy:
±
0.12 %
Frequency Range: 60 - 20000 Hz
Accuracy:
±
3 %
DC Voltage Range: 0 - 1000 V dc
Accuracy:
±
0.012 %
Current Range: 350
µ
A - 2 A
Accuracy: AC (60 Hz to 1 kHz):
±
0.25 %
DC:
±
0.05 %
Frequency Source: 19.999 kHz - 199.99 kHz
Accuracy:
±
0.0025 %
Amplitude: 150 mV to 6V rms
Accuracy:
±
5 %
Fluke 5500A Multi-Product
Calibrator or equivalent
Range:
1
Ω
- 100 M
Ω
Accuracy: 0.065 %
TC Adapter Accessory
K-type
Fluke 80 AK
K-type Thermocouple
K-type, mini-plug on both ends
Performance Tests
The following performance tests verify the complete operability of the Meter and check
the accuracy of each Meter function against the Meter’s specifications. Performance tests
should be performed annually to ensure that the Meter is within accuracy specifications.
Accuracy specifications are valid for a period of one year after calibration adjustment,
when measured at an operating temperature of 18
°
C to 28
°
C and at a maximum of 90 %
relative humidity.
To perform the following tests, it is not necessary to open the case. No adjustments are
necessary. Make the required connections, apply the designated inputs, and determine if
the reading on the Meter display falls within the acceptable range indicated.
Note
If the Meter fails any of these tests, it needs calibration adjustment or repair.