87 V/AN
Service Manual
6
Table 2. AC Voltage Function Specifications
Function
Range Resolution
Accuracy
45 - 65 Hz
30 - 200 Hz 200 - 440 Hz
440 Hz - 1
kHz
1 - 5 kHz
5 - 20 kHz
1
±
(0.7 % + 4)
±
(0.7 % + 2)
±
(2.0 % + 4)
±
(2.0 % + 20)
600.0 mV
6.000 V
60.00 V
600.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
±
(2.0 % + 4)
3
unspecified
1000 V
1 V
±
(1.0 % + 4)
unspecified unspecified
K
2,4
Low pass filter
±
(0.7 % + 2)
±
(1.0 % + 4)
+1 % + 4
-6 % - 4
5
unspecified
unspecified unspecified
1.
Below 10 % of range, add 12 counts.
2. The
Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter
may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit
change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error
in later measurements.
3.
Frequency range: 1 kHz to 2.5 kHz.
4.
A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
5.
Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.
Table 3. DC Voltage, Resistance, and Conductance Function Specifications
Function Range
Resolution
Accuracy
L
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
±
(0.05 % + 1)
±
(0.05 % + 1)
±
(0.05 % + 1)
±
(0.05 % + 1)
F
mV
600.0 mV
0.1 mV
±
(0.1 % + 1)
e
nS
600.0
Ω
6.000 k
Ω
60.00 k
Ω
600.0 k
Ω
6.000 M
Ω
50.00 M
Ω
60.00 nS
0.1
Ω
0.001 k
Ω
0.01 k
Ω
0.1 k
Ω
0.001 M
Ω
0.01 M
Ω
0.01 nS
±
(0.2 % + 2)
1
±
(0.2 % + 1)
±
(0.2 % + 1)
±
(0.6 % + 1)
±
(0.6 % + 1)
±
(1.0 % + 3)
2
±
(1.0 % + 10)
1
1.
When using the REL
∆
function to compensate for offsets.
2.
Add 0.5 % of reading when measuring above 30 M
Ω
in the 50 M
Ω
range.
Table 4. Temperature Specifications
Temperature Resolution Accuracy
1,2
-200
°
C to +1090
°
C
-328
°
F to +1994
°
F
0.1
°
C
0.1
°
F
1 % + 10
1 % + 18
1. Does not include error of the thermocouple probe.
2. Accuracy specification assumes ambient temperature stable to
±
1
°
C
. For ambient temperature changes of
±
5
°
C,
rated accuracy applies after 1 hour.