Fluke 79559999 Service Information Download Page 8

T5-600/T5-1000
Electrical Tester

8

Resistance and Continuity Function Tests

1.

 

Set the calibrator to the resistance given in Step 1 of Table 5.

2.

 

Apply the resistance to the tester. Verify that the tester reads within the display limits shown.

3.

 

Apply the resistances and verify the tester’s responses for the remaining steps in Table 5.

Table 5. Resistance and Continuity Tests

Step

Calibrator Output

T5-600/T5-1000 Display Limits and Beeper Responses

1

1  k

988 to 1012

2

26 

Beeper ON

3

300 

Beeper OFF

4

-2 to 2, Beeper ON

Low Battery Indicator Test

The following procedure verifies correct operation of the low battery indicator.

1.

 

Remove the tester’s batteries.

2.

 

Set the dc power supply to 3.0 V. Apply this voltage to the tester’s battery terminals.

3.

 

Set the calibrator to 1 k

. Apply this resistance to the tester’s probes.

4.

 

Slowly decrease the dc voltage just until the tester’s low battery indicator (

M

) turns on.

5.

 

Verify that the tester reads 1000 

±

12 counts (988 to 1012).

6.

 

Disconnect the calibrator and the dc power supply. Reinstall the tester’s batteries.

Parts and Accessories

Tables 6 shows the replacement parts and accessories available from Fluke for the T5-600 and T5-1000
Electrical Testers.

Table 6. Replacement Parts and Accessories

Description

Fluke Part Number

Test lead assembly

 Replace only with Fluke double-insulated leads.

648029

Battery door

1626588

Battery door screw

1618578

AA battery, 1.5 V, carbon-zinc (2 required)

650181

AA battery, 1.5 V, alkaline (2 required)

376756

T5-600/T5-1000 Electrical Tester Instruction Sheet (English)

1629509

Shop for Fluke products online at:

1.800.561.8187

www.

MyFlukeStore

.ca

Summary of Contents for 79559999

Page 1: ...n may be impaired Before each use Make sure the battery door is closed and latched Inspect the tester and test leads Look for cracks missing plastic exposed metal or damaged insulation Replace damaged...

Page 2: ...Category III Pollution Degree 2 T5 1000 1000 V rms Overvoltage Category III 600 V RMS OverVoltage Category IV Pollution Degree 2 Maximum Voltage Between Current Fork and Earth Ground T5 600 and T5 100...

Page 3: ...nt e 1 65 V dc nominal 600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel and lighting systems in large buildings CAT IV protects against transien...

Page 4: ...gits e 1000 1000 1 1 2 digits Input Characteristics Input Protection Function T5 600 T5 1000 Input impedance nomial K 600 V rms 1000 V rms 1 M 100 pF ac coupled L 600 V rms 1000 V rms 1 M 100 pF e 600...

Page 5: ...the coil cause the tester to read 10 times more current than the calibrator supplies 1 Put one edge of the butterfly coil in a vice to hold the coil during testing 2 Set the calibrator to the current...

Page 6: ...ingle conductor magnet wire Separate into two 5 turn coils Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f eps Figure 1 Making the Butterfly Coil Shop f...

Page 7: ...ator ON Calibrator Output T5 1000 Display Limits AC annunciator ON 600 V 60 Hz 589 to 611 1000 V 60 Hz 983 to 1017 Table 3 DC Voltage Tests T5 600 T5 1000 Step Calibrator Output T5 600 Display Limits...

Page 8: ...tage to the tester s battery terminals 3 Set the calibrator to 1 k Apply this resistance to the tester s probes 4 Slowly decrease the dc voltage just until the tester s low battery indicator M turns o...

Page 9: ...tion Sheet International 1621978 H5 Belt Holster Accessory TP1 Probe Set Flat Blade Accessory TP4 Probe Set 4 mm Round Accessory Battery Replacement Figure 2 shows how to replace the batteries Observe...

Reviews: