Fluke 79559999 Service Information Download Page 3

Service Information

Serial Numbers 79559999 and Below

3

General Specifications (continued)

RF Field Specification

0.5 % full scale + (specified accuracy) at 3 V/m

Safety

Complies with ANSI/ISA-S82.01-94, UL Classified to IEC 61010, CSA/CAN C22.2 No.1010.1-
92, and EN61010-1 1995.

EMC

EN 61326

Certifications

  

 

  

 

In accordance with IEC 61010-1 54CJ

Resolution and Accuracy

Function

T5-600 Range

T5-1000 Range

Resolution

Accuracy

K

600 V rms

1000 V rms

1 V

±

(1.5 % + 2 digits)

L

600 V

1000 V

1 V

±

(1 % + 1 digit)

?

100.0 A

100.0 A

0.1 A

±

(3 % + 3 digits)

e

1000 

1000 

±

(1 % + 2 digits)

Input Characteristics

Input Protection

Function

T5-600

T5-1000

Input impedance (nomial

K

600 V rms

1000 V rms

1 M

, <100 pF ac-coupled

L

600 V rms

1000 V rms

1 M

, <100 pF

e

600 V rms

1000 V rms

Open Circuit Test Voltage

Short Circuit Current

e

1.65 V dc (nominal)

<600 

µ

A

CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in
large buildings.

CAT IV protects against transients from a primary supply such as an electricity meter or an overhead or underground utility
service.

Serial Numbers 79559999 and Below

General Specifications

Calibration

One-year calibration cycle.

Maximum Voltage
Between any Terminal
and Earth Ground

T5-600:  600 V rms, Overvoltage  Category III
T5-1000:  1000 V rms, Overvoltage Category III

Temperature

Operating:  -10 

°

C to +50 

°

C (14 

°

F to 122 

°

F); Storage:  -30 

°

C to +60 

°

C (-22 

°

F to +140 

°

F)

Altitude

Operating:  2000 m (6562 ft); Storage: 10,000 m (32808 ft)

Relative Humidity

0 % to 95 %, 5 

°

C to 30 

°

C (41 

°

F to 86

°

F); 0 % to 75 %, 30 

°

C to 40 

°

C (86 

°

F to 104 

°

F);

0 % to 45 %, 40 

°

C to 50 

°

C (104 

°

F to 122 

°

F)

Battery Type and Life

AA (2); 400 hours continuous with alkaline; 200 hours continuous with zinc chloride

Shop for Fluke products online at:

1.800.561.8187

www.

MyFlukeStore

.ca

Summary of Contents for 79559999

Page 1: ...n may be impaired Before each use Make sure the battery door is closed and latched Inspect the tester and test leads Look for cracks missing plastic exposed metal or damaged insulation Replace damaged...

Page 2: ...Category III Pollution Degree 2 T5 1000 1000 V rms Overvoltage Category III 600 V RMS OverVoltage Category IV Pollution Degree 2 Maximum Voltage Between Current Fork and Earth Ground T5 600 and T5 100...

Page 3: ...nt e 1 65 V dc nominal 600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel and lighting systems in large buildings CAT IV protects against transien...

Page 4: ...gits e 1000 1000 1 1 2 digits Input Characteristics Input Protection Function T5 600 T5 1000 Input impedance nomial K 600 V rms 1000 V rms 1 M 100 pF ac coupled L 600 V rms 1000 V rms 1 M 100 pF e 600...

Page 5: ...the coil cause the tester to read 10 times more current than the calibrator supplies 1 Put one edge of the butterfly coil in a vice to hold the coil during testing 2 Set the calibrator to the current...

Page 6: ...ingle conductor magnet wire Separate into two 5 turn coils Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f eps Figure 1 Making the Butterfly Coil Shop f...

Page 7: ...ator ON Calibrator Output T5 1000 Display Limits AC annunciator ON 600 V 60 Hz 589 to 611 1000 V 60 Hz 983 to 1017 Table 3 DC Voltage Tests T5 600 T5 1000 Step Calibrator Output T5 600 Display Limits...

Page 8: ...tage to the tester s battery terminals 3 Set the calibrator to 1 k Apply this resistance to the tester s probes 4 Slowly decrease the dc voltage just until the tester s low battery indicator M turns o...

Page 9: ...tion Sheet International 1621978 H5 Belt Holster Accessory TP1 Probe Set Flat Blade Accessory TP4 Probe Set 4 mm Round Accessory Battery Replacement Figure 2 shows how to replace the batteries Observe...

Reviews: