Fluke 79559999 Service Information Download Page 2

T5-600/T5-1000
Electrical Tester

2

Parts and Service

The tester is warranted to be free from defects in material and workmanship for two years, while under normal
use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the
T5-600/T5-1000 Electrical Tester Instruction Sheet.

Specifications

Accuracy is specified for one year after calibration, at 18 

°

C to 28 

°

C (64 

°

F to 82 

°

F) with relative humidity

to 90 %. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave
input. Accuracy specifications are given as follows:

±

([ % of reading] + [number of least significant digits])

Temperature coefficient of 0.1 x (specified accuracy)/ 

°

C for <18 

°

C or >28 

°

C (<64.4 

°

F or >82.4 

°

F)

 

 

 

Serial Numbers 79560000 and Above

General Specifications

Calibration

One-year calibration cycle.

Maximum Voltage
Between any Terminal
and Earth Ground

T5-600:  600 V rms, Overvoltage  Category III, Pollution Degree 2
T5-1000:  1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV,
Pollution Degree 2

Maximum Voltage
Between Current Fork
and Earth Ground

T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage
Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.)

Temperature

Operating:  -10 

°

C to +50 

°

C (14 

°

F to 122 

°

F); Storage:  -30 

°

C to +60 

°

C (-22 

°

F to +140 

°

F)

Altitude

Operating:  2000 m (6562 ft); Storage: 10,000 m (32808 ft)

Relative Humidity

0 % to 95 %, 5 

°

C to 30 

°

C (41 

°

F to 86

°

F); 0 % to 75 %, 30 

°

C to 40 

°

C (86 

°

F to 104 

°

F);

0 % to 45 %, 40 

°

C to 50 

°

C (104 

°

F to 122 

°

F)

Battery Type and Life

AA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chloride

Shock, Vibration

1 m drop at -10 

°

C to + 50 

°

C (14 

°

F to 122 

°

F) per ANSI/ISA-S82.01-1994 and EN 61010-1

1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)

Surge Protection

T5-600:  6 kV per IEC 61010, T5-1000:  8 kV per IEC 61010

Enclosure Rating

IP 52 per IEC 60529, no vacuum applied

Shop for Fluke products online at:

1.800.561.8187

www.

MyFlukeStore

.ca

Summary of Contents for 79559999

Page 1: ...n may be impaired Before each use Make sure the battery door is closed and latched Inspect the tester and test leads Look for cracks missing plastic exposed metal or damaged insulation Replace damaged...

Page 2: ...Category III Pollution Degree 2 T5 1000 1000 V rms Overvoltage Category III 600 V RMS OverVoltage Category IV Pollution Degree 2 Maximum Voltage Between Current Fork and Earth Ground T5 600 and T5 100...

Page 3: ...nt e 1 65 V dc nominal 600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel and lighting systems in large buildings CAT IV protects against transien...

Page 4: ...gits e 1000 1000 1 1 2 digits Input Characteristics Input Protection Function T5 600 T5 1000 Input impedance nomial K 600 V rms 1000 V rms 1 M 100 pF ac coupled L 600 V rms 1000 V rms 1 M 100 pF e 600...

Page 5: ...the coil cause the tester to read 10 times more current than the calibrator supplies 1 Put one edge of the butterfly coil in a vice to hold the coil during testing 2 Set the calibrator to the current...

Page 6: ...ingle conductor magnet wire Separate into two 5 turn coils Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f eps Figure 1 Making the Butterfly Coil Shop f...

Page 7: ...ator ON Calibrator Output T5 1000 Display Limits AC annunciator ON 600 V 60 Hz 589 to 611 1000 V 60 Hz 983 to 1017 Table 3 DC Voltage Tests T5 600 T5 1000 Step Calibrator Output T5 600 Display Limits...

Page 8: ...tage to the tester s battery terminals 3 Set the calibrator to 1 k Apply this resistance to the tester s probes 4 Slowly decrease the dc voltage just until the tester s low battery indicator M turns o...

Page 9: ...tion Sheet International 1621978 H5 Belt Holster Accessory TP1 Probe Set Flat Blade Accessory TP4 Probe Set 4 mm Round Accessory Battery Replacement Figure 2 shows how to replace the batteries Observe...

Reviews: