Fluke 79559999 Service Information Download Page 4

T5-600/T5-1000
Electrical Tester

4

General Specifications (continued)

Shock, Vibration

1 m drop at 15 

°

C to 35 

°

C (59 

°

F to 95 

°

F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1993.

Sinusoidal vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)

Surge Protection

T5-600:  6 kV per IEC 1010-1, 1990-09: T5-1000:  8 kV per IEC 1010-1, 1990-09

Enclosure Rating

IP 52 per IEC 60529, no vacuum applied

RF Field Specification

0.5 % full scale + (specified accuracy) at 3 V/m

Safety

Complies with ANSI/ISA-S82.01-94 for use in overvoltage category III (CAT III) locations,
UL3111, CSA/CAN C22.2 No.1010.1-92, and EN61010-1 1993.

EMC

EN 50081-1, EN 50082-1

Certifications

 

 

 

Resolution and Accuracy

Function

T5-600 Range

T5-1000 Range

Resolution

Accuracy

K

600 V rms

1000 V rms

1 V

±

(1.5 % + 2 digits)

L

600 V

1000 V

1 V

±

(1 % + 1 digit)

?

100.0 A

100.0 A

0.1 A

±

(3 % + 3 digits)

e

1000 

1000 

±

(1 % + 2 digits)

Input Characteristics

Input Protection

Function

T5-600

T5-1000

Input impedance (nomial

K

600 V rms

1000 V rms

1 M

, <100 pF ac-coupled

L

600 V rms

1000 V rms

1 M

, <100 pF

e

600 V rms

1000 V rms

Open Circuit Test Voltage

Short Circuit Current

e

2.4 V dc (nominal)

<600 

µ

A

CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in
large buildings.

CAT IV protects against transients from a  primary supply such as an electricity meter or an overhead or underground utility
service.

Cleaning the Tester

W

Warning

To avoid electric shock or damage to the tester, never allow water inside the
case. To avoid damaging the tester’s case, never use solvents on the tester.

If the tester requires cleaning, wipe it down with a cloth that is lightly dampened with water or a mild
detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-based fluids when wiping
down the tester.

Shop for Fluke products online at:

1.800.561.8187

www.

MyFlukeStore

.ca

Summary of Contents for 79559999

Page 1: ...n may be impaired Before each use Make sure the battery door is closed and latched Inspect the tester and test leads Look for cracks missing plastic exposed metal or damaged insulation Replace damaged...

Page 2: ...Category III Pollution Degree 2 T5 1000 1000 V rms Overvoltage Category III 600 V RMS OverVoltage Category IV Pollution Degree 2 Maximum Voltage Between Current Fork and Earth Ground T5 600 and T5 100...

Page 3: ...nt e 1 65 V dc nominal 600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel and lighting systems in large buildings CAT IV protects against transien...

Page 4: ...gits e 1000 1000 1 1 2 digits Input Characteristics Input Protection Function T5 600 T5 1000 Input impedance nomial K 600 V rms 1000 V rms 1 M 100 pF ac coupled L 600 V rms 1000 V rms 1 M 100 pF e 600...

Page 5: ...the coil cause the tester to read 10 times more current than the calibrator supplies 1 Put one edge of the butterfly coil in a vice to hold the coil during testing 2 Set the calibrator to the current...

Page 6: ...ingle conductor magnet wire Separate into two 5 turn coils Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f eps Figure 1 Making the Butterfly Coil Shop f...

Page 7: ...ator ON Calibrator Output T5 1000 Display Limits AC annunciator ON 600 V 60 Hz 589 to 611 1000 V 60 Hz 983 to 1017 Table 3 DC Voltage Tests T5 600 T5 1000 Step Calibrator Output T5 600 Display Limits...

Page 8: ...tage to the tester s battery terminals 3 Set the calibrator to 1 k Apply this resistance to the tester s probes 4 Slowly decrease the dc voltage just until the tester s low battery indicator M turns o...

Page 9: ...tion Sheet International 1621978 H5 Belt Holster Accessory TP1 Probe Set Flat Blade Accessory TP4 Probe Set 4 mm Round Accessory Battery Replacement Figure 2 shows how to replace the batteries Observe...

Reviews: