74X Series
Calibration Manual
10
General Specifications
Display:
240 by 200 pixel graphic LCD, 70 x 58 mm.
Power:
Internal battery pack: NiCd, 7.2V, 1700 mAh.
Memory Backup:
Lithium battery, 5 years typical lifetime.
Dimensions:
130 x 236 x 61 mm (5.1 x 9.3 x 2.4 in.).
Weight:
1.4 kg (3 lb. 1 oz.).
Altitude:
Up to 2800 meters (9186 ft) above mean sea level.
Operating
-10 to 50
°
C (typically to -20
°
C, except for frequency measure and
Temperature:
ac voltage measure).
Storage Temperature:
-20 to 60
°
C
Humidity:
Avoid prolonged use outside the safe operating boundaries shown in the
graph on the next page.
RF Fields:
Accuracy for all functions is not specified in RF fields >3 V/m
Accuracy for thermocouple measurement is not specified in RF fields >1
V/m
Accuracy for ohms/RTD source is not specified in RF fields >0.5 V/m
Accuracy for mADC measurement is not specified in RF fields >1.5 V/m
Safety:
Designed in accordance with CAT II 300 Volts Pollution Degree 2, IEC
1010-1, ANSI/ISA-S82, UL3111, and CSA C22.2 No. 1010.1-92. See
“Safety Information” near the front of this manual.
Warranty:
See the WARRANTY, inside front cover of the Users Manual.
100
90
80
70
60
50
40
30
20
10
0
-20
0
20
40
60
80
100
120
140
Temperature (
˚
F)
Temperature (
˚
C)
%RH
-4
-20
0
30
40
50
60
= Normal Operation (-10
˚
C — 50
˚
C)
= Storage (-20
˚
C — 60
˚
C)
+
14
-10
gb02f.eps
Figure 1. LCD Operating Environment Specification