Performance Verification Tests
21
Proceed as follows to verify the Resistance Source function.
1.
Connect the UUT to the HP3458A as shown in Figure 9. Use a four-wire connection
transitioning to two wires at the UUT.
2.
Set the UUT to the Resistance Source function at 0.1
Ω
.
3.
On the 3458A, select four-wire ohms measurement and up-range to the 100
Ω
range.
Use the 100
Ω
range for the first three tests points, and autorange thereafter. The low
range of the 3458A supplies too much current back into the UUT.
4.
Observe that the reading on the 3458A is within the range shown under the
appropriate column in Table 11.
5.
Continue through the test points, verifying compliance to specifications.
6.
When you are finished, press
c
twice on the UUT to set the Source function to off.
This saves battery life.
TC
mA
mA
V
V
RTD
RTD
SOURCE
300V
MAX
30V
MAX
30V
MAX
30V
MAX
MEAS
CAT
FLUKE 74X
Ω
SENSE
INPUT
HI
LO
GUARD
HI
LO
HP3458A MULTIMETER
gb10f.eps
Figure 9. Resistance Source Verification Connections
Table 11. Resistance Source Verification Points
UUT
Range
UUT
Output
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
11 ohm
0.1 ohm
0.0800
0.1200
0.0800
0.1200
11 ohm
1 ohm
0.9799
1.0201
0.9798
1.0202
11 ohm
10 ohm
9.9790
10.0210
9.9780
10.0220
110 ohm
20 ohm
19.958
20.042
19.956
20.044
110 ohm
100 ohm
99.950
100.050
99.940
100.060
1100 ohm
200 ohm
199.46
200.54
199.44
200.56
1100 ohm
1000 ohm
999.30
1000.70
999.20
1000.80
11 kohm
2 kohm
1.9944
2.0056
1.9942
2.0058
11 kohm
10 kohm
9.9920
10.0080
9.9910
10.0090