Press the “SET” button to set the BAUD rate to the new value or the “EXIT”
button to abort the operation and skip to the next parameter in the menu.
9.12.2
Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure-
ments transmitted from the serial interface. If the sample rate is set to 5, the
bath transmits the current measurement over the serial interface approximately
every five seconds. The automatic sampling is disabled with a sample period of
0. The sample period is indicated by,
SAMPLE
Serial sample period parameter
Press the “SET” button to choose to set the sample period. The current sample
period value is displayed.
SA = 1
Current sample period (seconds)
Adjust the value with the “UP” or “DOWN” button and then use the “SET”
button to set the sample rate to the displayed value.
SA = 60
New sample period
9.12.3
Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du-
plex or half duplex. With full duplex any commands received by the bath via
the serial interface are immediately echoed or transmitted back to the device of
origin. With half duplex the commands are executed but not echoed. The du-
plex mode parameter is indicated by,
DUPL
Serial duplex mode parameter
Press the “SET” button to access the mode setting
DUP=FULL
Current duplex mode setting
The mode may be changed using the “UP” or “DOWN” button and then press-
ing the “SET” button.
DUP=HALF
New duplex mode setting
9.12.4
Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
53
9 Controller Operation
Serial Interface Parameters