![Fluke 7312 User Manual Download Page 19](http://html.mh-extra.com/html/fluke/7312/7312_user-manual_2309393019.webp)
4
Quick Start
CAUTION:
Read Section 6 entitled BATH USE before placing the bath in
service.
This chapter gives a brief summary of the steps required to set up and operate
the bath. This should be used as a general overview and reference and not as a
substitute for the remainder of the manual. Please read Section 6, Installation,
through Section 9, General Operation, carefully before operating the bath.
4.1
Unpacking
Unpack the bath carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier immedi-
ately. Verify that all components are present:
•
7312 Bath
•
Access Hole Cover and Components.
•
Triple Point of Water Support Racks and Adapters.
•
User’s Guide
•
Report of Test
•
Drain Valve
If you are missing any item, please call an Authorized Service Center for
assistance.
4.2
Set Up
CAUTION:
The drain valve must be installed on the back of the bath be-
fore attempting to fill the tank with fluid. See Section 5.3, on page 17 for
drain installation instructions.
WARNING:
The instrument is equipped with a soft cutout (user settable
firmware) and a hard cutout (set at the factory). Check the flash point,
boiling point, or other fluid characteristic applicable to the circumstances
of the unit operation. Ensure that the soft cutout is adjusted to the fluid
characteristics of the application. As a guideline, the soft cutout should be
set 10°C to 15°C below the flash point of the bath fluid. See Section 8.1,
Heat Transfer Fluid, for specific information on bath fluids and Section
9.8, Cutout.
Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, and connecting power. Consult Section 6, Installation, for
13
4 Quick Start
Unpacking
Summary of Contents for 7312
Page 2: ......
Page 3: ......
Page 18: ......
Page 29: ...25 6 Bath Use Triple Point of Water Cell Maintenance Fluid Level in Bath Figure 3 Fluid Level ...