detailed instructions for proper installation of the bath. Be sure to place the
bath in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at moderate bath
temperatures, clean distilled water works well above 0°C and below 70°C.
Carefully pour the fluid into the bath tank through the large rectangular access
hole above the tank avoiding spilling any fluid. The fluid must not exceed a
height of 12.7-20.3 mm (0.5-0.8 inches) below the bottom of the lid (
NOT
the
access cover).
4.3
Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency
and current capability. Refer to Section 3.1, Specifications, for power details.
Refer to and read the caution at the front of the manual concerning brownout
and over voltage protection. Check the back panel label for the correct voltage
and frequency prior to energizing the unit. Turn the bath on using the front
panel “POWER” switch. The baths turns on and begins to heat or cool to reach
the previously programmed temperature set-point. The front panel LED display
indicates the actual bath temperature. Set the cooling switch to “OFF” for tem-
perature above approximately 45°C. Set the switch to “ON” for lower
temperatures.
4.4
Setting the Temperature
In the following discussion and throughout this manual a solid box around the
word SET, UP, DOWN, or EXIT. Indicates the panel button to press while the
dotted box indicates the display reading on the front panel. Explanation of the
button function or display reading is written at the right.
To view or set the bath temperature set-point proceed as follows. The front
panel LED display normally shows the actual bath temperature.
24.68 C
Bath temperature display
When the “SET” button is pressed the display shows the set-point memory that
is currently being used and its value. Eight set-point memories are available.
S
Access set-point selection.
Set-point 1, 25.0°C currently used
Press the “SET” button to select this memory and access the set-point value.
S
Access set-point value
C 25.00
Current value of set-point 1, 25.00°C
Press the “UP” or “DOWN” button to change the set-point value.
7312 TPW Maintenance Bath
User’s Guide
14
Summary of Contents for 7312
Page 2: ......
Page 3: ......
Page 18: ......
Page 29: ...25 6 Bath Use Triple Point of Water Cell Maintenance Fluid Level in Bath Figure 3 Fluid Level ...