again. If the mode is set to manual, then the heater will remain disabled until
the user manually resets the cutout.
The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.
Access heater power in secondary menu
12 Pct
Heater power in percent
Access proportional band
Pb=0.101C
Proportional band setting
Access cutout set-point
CO= 110C
Cutout set-point
To change the cutout set-point press “UP” or “DOWN”.
CO= 75C
New cutout set-point
To accept the new cutout set-point press “SET”.
Accept cutout set-point
The next function is the configuration menu. Press “EXIT” to resume display-
ing the bath temperature.
9.9
Controller Configuration
The controller has a number of configuration and operating options and calibra-
tion parameters which are programmable via the front panel. These are ac-
cessed from the secondary menu after the cutout set-point function by pressing
“SET.” There are 5 sets of configuration parameters — probe parameters, oper-
ating parameters, serial interface parameters, IEEE-488 interface parameters,
and controller calibration parameters. The menus are selected using the “UP”
and “DOWN” keys and then pressing “SET”. See Figure 6.
9.10
Probe Parameters Menu
The probe parameter menu is indicated by,
PrObE
Probe parameters menu
37
Controller Operation