S
Access set-point selection
1. 25.0
Set-point 1, 25.0°C currently used
Press “SET” to select this memory and access the set-point value.
S
Access set-point value
C 25.00
Current value of set-point 1, 25.00°C
Press “UP” or “DOWN” to change the set-point value.
U
Increment display
C 30.00
New set-point value
Press SET to accept the new value and display the vernier value. The bath be-
gins heating or cooling to the new set-point.
Store new set-point, access vernier
0.00000
Current vernier value
Press “EXIT” and the bath temperature will be displayed again.
Return to the temperature display
24.73 C
Bath temperature display
The bath will heat or cool until it reaches the new set-point temperature. Set the
heater switch to position “HIGH” to allow the bath to more quickly reach a
higher temperature. The “HIGH” setting may be necessary to reach higher tem-
peratures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set to
prevent this from happening. See Section 8.8.
If operating the bath below 45 °C set the COOLING power switch to ON. The
cooling temperature may require adjustment to provide the proper amount of
cooling. See section 8.5.
To obtain optimum control stability adjust the proportional band as discussed in
Section 9.7.
User’s Guide
12