COLD BATHS
•
Refrigerated baths require that the condensing coil be cleaned periodi-
cally. Accumulation of dust and dirt on the condenser will result in pre-
mature failure of the compressor.
•
This bath has been equipped with a brownout and over voltage protection
device as a safety feature to protect the system components.
•
Mode of Operation:
This bath needs to be plugged into the line voltage
for at least 2 minutes before operation. This is only necessary for the first
time the bath is energized or when it is moved from one location to an-
other. Turning the bath ON or OFF does not trigger the delay.
•
If a High/Low voltage condition exists for longer than 5 seconds, the bath
de-energizes. An amber indicator on the back panel lights when this con-
dition exists.
•
Re-energization is automatic upon correction of the fault condition and af-
ter a delay cycle of about 2 minutes. If a fault condition exists upon appli-
cation of power, the bath will not energize.
•
Under and Over Voltage Protection at 230 VAC
♦
Voltage Cutout: ±12.5% (203 - 257 VAC)
♦
Voltage Cut In: ±7.5% (213 - 247 VAC)
1.3