
more than 10 days, the instrument needs to be energized for a "dry-out"
period of 2 hours before it can be assumed to meet all of the safety re-
quirements of the IEC 1010-1. If the product is wet or has been in a wet
environment, take necessary measures to remove moisture prior to apply-
ing power such as storage in a low humidity temperature chamber operat-
ing at 50°C for 4 hours or more.
•
Overhead clearance is required. Do not place the instrument under a cabi-
net or other structure. Always leave enough clearance to allow for safe
and easy insertion and removal of probes.
•
The instrument is intended for indoor use only.
•
The bath is a precision instrument. Although it has been designed for op-
timum durability and trouble free operation, it must be handled with care.
Position the bath before the tank is filled with fluid by rolling it into
place.
DO NOT
attempt to lift the bath.
DO NOT move a bath filled
with fluid.
BURN HAZARD
•
Extremely cold temperatures may be present in this equipment. Freezer
burns and frostbite may result if personnel fail to observe safety precau-
tions.
•
High temperatures may be present in this equipment. Fires and severe
burns may result if personnel fail to observe safety precautions.
ELECTRICAL HAZARD
•
These guidelines must be followed to ensure that the safety mechanisms
in this instrument will operate properly. This instrument must be plugged
into a 230V AC electric outlet of the appropriate frequency. The power
cord of the instrument is equipped with a three-pronged grounding plug
for your protection against electrical shock hazards. It must be plugged
directly into a properly grounded three-prong receptacle. The receptacle
must be installed in accordance with local codes and ordinances. Consult
a qualified electrician. DO NOT use an extension cord or adapter plug.
•
DO
use a ground fault interrupt device. This instrument contains a fluid.
A ground fault device is advised in case fluid is present in the electrical
system and could cause an electrical shock.
•
Always replace the power cord with an approved cord of the correct rat-
ing and type. If you have questions, contact a Fluke Fluke Calibration Au-
thorized Service Center (see Section 1.3).
•
High voltage is used in the operation of this equipment. Severe injury or
death may result if personnel fail to observe the safety precautions. Before
working inside the equipment, turn off the power and disconnect the
power cord.
3
Before You Start