562/563
Users Manual
20
If you cannot paint or use tape, then you can improve the
accuracy of your measurements by either numerically adjusting
emissivity or by using the Emissivity Menu to access a table of
some common materials. The Product has numerically
adjustable emissivity from 0.10 to 1.00 that can be used in
conjunction with the emissivity in Table . Also, the Product has a
built-in table of the most common materials that can be set from
the Emissivity menu. The built-in values are indicated in Table .
HOLD
After the trigger is released, the display retains its last infrared
measurement for 20 seconds. At the same time,
HOLD
shows
on the display. With the probe inserted, the Product stays on. To
freeze the infrared temperature when a probe is not inserted,
release the trigger until
HOLD
shows on the display.
Data Storage
The 562 can store up to 20 data records. The 563 can store up
to 99 data records. Each record includes:
•
Record
number
•
IR and probe temperature in
°
F or
°
C
•
Date/Time
•
Emissivity
•
Max/Min/Avg/Dif Temperature (if enabled)
For more information, see “Save”.