background image

5128A 

Operators Manual 

24 

For short probes without an available cable, insert the probe into the door as far 
as possible and do calibrations at or near ambient temperature. This eliminates 
temperature gradients between the UUT and the chamber. Under these 
conditions, where the UUT probe is not fully immersed in the working volume, an 
external reference probe that is inserted at the same depth should be used to 
ensure the best calibration. 

Stabilize the Chamber and UUT 

The Product response time is normally much faster than that of the UUT, with 
typical stabilization times <5 minutes between typical UUT verification points. 
However, to ensure that the UUT temperature and humidity is at equilibrium with 
the chamber conditions, allow for stabilization times of at least 15 minutes 
between set points. Longer stabilization times may be necessary if the initial UUT 
temperature is significantly different than the chamber temperature and if the 
chamber is set to extreme humidity and temperature set points. Monitor the 
actual temperature and humidity points as shown on the Product display. 
Chamber stabilization is typically reached when the actual temperature is within 
0.1 

°

C of the set temperature, and the actual humidity is within 0.2 % of the set 

humidity. 

Data Logger Calibration 

To calibrate Data Loggers: 

1.  Place them directly into the calibration area of the chamber and use the clear 

glass door and the shelf accessory. See Figure 4 for approximate placement. 

 

ibw025.eps 

Figure 4. Data Logger Placement 

2.  Put the UUT into its data logging mode, and then set and record the actual 

calibrator temperature and humidity at known points in time.  

3.  Compare the UUT logged data against the actual Product temperature and 

humidity settings.

 

For data loggers that are too large to fit inside the mixing 

insert, but fit inside the outer chamber, the larger shelf can be used to hold 
the UUT. Use an external reference probe to monitor the conditions of the 
Product. 

Summary of Contents for 5128A

Page 1: ...cifications are subject to change without notice All product names are trademarks of their respective companies 5128A RHapid Cal Humidity Generator Operators Manual Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it ...

Page 2: ... nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepaid FOB Destination If Fluke determines that failure was caus...

Page 3: ...up 8 Unpack and Inspect 8 Product Placement 8 Fill the Product with Distilled Water 9 Mains Voltage 10 The Product 12 Startup and Main Screen 15 Maintenance Screens 15 Probe Calibrate Screen 16 Probe Calibrate Screen 16 BIT Humidity Gen Screen 16 Error Code Screen 16 Firmware Version Screen 16 Operation 18 Chamber Door Use 18 Door Styles 19 Turn On the Product 19 Turn the Product Off 20 Mixing Ins...

Page 4: ...ip the Product 27 Maintenance 27 Clean the Exterior 27 Desiccant Replacement 28 Water Level 29 Fuse Replacement 29 Calibration 32 Calibration Principle 32 Environmental Conditions 32 Equipment Setup 32 Using a Chilled Mirror Probe 33 As Found Data Collection 36 Load Temperature and Humidity Offsets into the Product 37 As Left Data Collection 37 Temperature Uniformity Verification 38 Error Messages...

Page 5: ... humidity in the chamber The Product Controls and maintains the relative humidity and temperature in the chamber to within a resolution of 0 1 RH and 0 1 C respectively Has extremely fast response times for changes in humidity and temperature with ramp times as fast as 10 RH minute and 10 C minute Has different door styles to accommodate a variety of humidity sensors Features a simple front panel ...

Page 6: ...on s website at www flukecal com To register your product visit http flukecal com register product Safety Information A Warning identifies conditions and procedures that are dangerous to the user A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test Warning To prevent possible electrical shock fire or personal injury Read all safety informa...

Page 7: ...Product can be compromised Do not put the Product where access to the mains power cord is blocked Do not use the Product if it is damaged Disable the Product if it is damaged Do not use the Product if it operates incorrectly Do not operate the Product with covers removed or the case open Hazardous voltage exposure is possible Use only specified replacement parts Use only specified replacement fuse...

Page 8: ...ve marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE Directive Annex I this product is classed as category 9 Monitoring and Control Instrumentation product Do not dispose of this product as unsorted municipal waste Calibration and Repair Informa...

Page 9: ...e directly connected to a low voltage power supply network that supplies buildings used for domestic purposes There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such envir...

Page 10: ...is the maximum humidity value at which the specification applies See Chamber Operational Range below for Hmax Note Specifications apply to the Working Volume shown in the Working Volume Template found at the end of the Operators Manual and are referenced to the Actual reading on the Product display Chamber Uniformity and Stability Ambient Temperature Range 23 C 3 C 1 Chamber Temperature Chamber Hu...

Page 11: ...d on temperature and humidity settings The table below shows the maximum RH attainable at different temperatures Operational Specifications Temperature rate of change down 1 5 C minute Typical Temperature rate of change up 10 C minute Typical Humidity rate of change down 5 RH minute Typical Humidity rate of change up 10 RH minute Typical ...

Page 12: ...libration and calibration label 1 Fill syringe with extension tube 1 Five port square chamber door 1 installed on the Product Port plug 5 installed on each probe port Silicone Unit Under Test UUT grommets in in in 1 in One of each size 4 in total Desiccant cartridge 1 installed inside the Product USB Cable Type A to Type B 1 Product Placement Put the Product on a clean flat surface For best result...

Page 13: ...o the Product with the plunger or bring the syringe to a higher level than the fill inlet and allow the water to flow into the Product Watch the water level indicator to gauge the fill level as the reservoir is filled The Product takes approximately two syringes of water Monitor the water level indicator to ensure the water level is between the Max and Min lines Caution To avoid damage to the Prod...

Page 14: ...onductor in the mains power cord is connected to a protective earth ground Disruption of the protective earth could put voltage on the chassis that could cause death Replace the mains power cord if the insulation is damaged or if the insulation shows signs of wear The Product requires mains voltage that ranges from 100 V ac to 240 V ac with frequencies from 50 Hz to 60 Hz The Product comes with th...

Page 15: ... Table 3 Available Mains Power Plug Types LC 1 LC 3 LC 4 LC 5 LC 6 LC 7 LC 42 hwr039 eps Type Fluke Calibration Option Number North America LC 1 Universal Euro LC 3 United Kingdom LC 4 Switzerland LC 5 Australia LC 6 South Africa LC 7 Brazil LC 42 ...

Page 16: ...le In Out Draws and returns gas from the Product chamber Use to measure the dew point of the chamber with a chilled mirror instrument that has its own pump The Sample In Out ports are the push to connect type and accept in OD tubing Fill Inlet and Water Level Indicator Use the fill inlet to fill the internal humidifier with distilled water The water level indicator shows the relative amount of wat...

Page 17: ...er the Product specifications See the end of this manual for a template that outlines the working area One of the three tabs on the outer part of the insert must be in the 6 o clock position as shown in the illustration This position ensures the best temperature and humidity uniformity within the working area Outer Chamber The chamber where the mixing insert is placed The UUT can be placed directl...

Page 18: ...old down NEXT for 4 seconds to step to the Standby screen 3 When in the Standby screen either down or up toggles the Standby mode On or Off Push and Hold EXIT down for 4 seconds to return to the main screen Down Button Moves the temperature or humidity setting downward depending on which parameter blinks A single button push increments the setting by one count Hold the button down to rapidly advan...

Page 19: ...number on the startup screen The Product then changes to the main screen and shows This indicates that the internal humidifier is warming up to create sufficient water vapor that will allow high relative humidity value settings The Product will not control the chamber temperature or humidity when is on the main screen To bypass this warmup period and control the chamber push EXIT once he Product n...

Page 20: ...An example of the calibration offset screen BIT Humidity Gen Screen The BIT humidity gen screen contains parameters associated with the internal humidity generator and is only used by trained service personnel Push and hold down EXIT to return to the main screen Error Code Screen The error code screen is shown and errors that the Product detects are presented See Error Codes for a list of error co...

Page 21: ...5 8 6 1 10 9 7 ibw003 eps Item Name Description Fan inlet outlet holes Do not block the inlet outlet holes Fans The two fans provide internal cooling to the Product Power supply fan access holes Provides cooling for the internal power supply Do not block these holes and periodically vacuum to remove any dust ...

Page 22: ...water from the Product Drain water if overfilled or for shipment HI HI terminal for the 12 V dc output to power up transmitters and transducers Fuse 12 V dc output fuse F1A 250 V fuse Operation The subsequent sections explain Product operation Before Product use see Fill the Product with Distilled Water Chamber Door Use Note Before the Product is powered on make sure that there is a chamber door s...

Page 23: ... the cap and placing the UUT through the cap and appropriate sized silicone grommet and port opening Accessory Clear square door with no ports no silicone gasket and four fasteners one at each corner To install and use this door use the four corner screws 1 4 in x 20 threads to fasten the door in place and tighten the door against the black outer seal on the outside of the chamber Use this door to...

Page 24: ...e chamber One of the three tabs on the outer part of the insert must be in the 6 o clock position as shown in Table 5 This position ensures the best temperature and humidity uniformity within the working area To check for water condensation remove the mixing insert from the chamber and then reinstall it Initial Warmup Period The Product is a fast response system able to change temperature as quick...

Page 25: ...level For example a UUT has verification points at 23 C 90 RH followed by 10 C 90 RH The first set point 23 C and 90 RH has a dew point of 21 3 within 2 C of the actual temperature When the chamber actual conditions reach this set point CAUTION shows on the display Before setting the chamber to 10 C set the humidity to about 1 2 the Actual 45 When that humidity is reached then set the Product to 1...

Page 26: ... or near the chamber door or on the UUT If condensation forms anywhere in the chamber the relative humidity inside the chamber is unknown Condensation must always be removed from inside the chamber before calibrations are performed See Remove Condensation Remove Condensation To remove condensation 1 Turn off the Product unplug the mains power cord and remove the chamber door 2 Carefully inspect th...

Page 27: ... probe inside the port opening Ensure the UUT sensor is inside the working volume of the chamber The working volume is outlined in Figure 11 and has a depth of 76 mm to 152 mm 3 in to 6 in from the inner door face 5 Hand tighten the cap around the UUT to give an airtight seal If necessary add sealant material around the UUT probe Parafilm Scotch Removable Adhesive Putty or similar materials work w...

Page 28: ...is significantly different than the chamber temperature and if the chamber is set to extreme humidity and temperature set points Monitor the actual temperature and humidity points as shown on the Product display Chamber stabilization is typically reached when the actual temperature is within 0 1 C of the set temperature and the actual humidity is within 0 2 of the set humidity Data Logger Calibrat...

Page 29: ... to avoid condensation on the Sample In Out port lines otherwise the lines must be heated to 5 C above the chamber temperature Fluke Calibration recommends that the gas sample is returned back to the Product through the Sample In port Measure the temperature within the chamber and the dew point measurement from the chilled mirror to determine the relative humidity inside the Product chamber For th...

Page 30: ...ommands Command Read Write Function BOOT Write Resets the Product HELP Read Returns the list of remote commands available ID Reads Returns Fluke 5128A serial number and firmware revision HOFFSET Read Returns the value of the humidity offset stored within the Product HOFFSET x x Write Sets the humidity offset used in the Product as part of the calibration process RH TOFFSET Read Returns the value o...

Page 31: ...ain the water by slightly tilting the Product When shipping the Product use the original carton and packing material or Fluke Calibration case accessory part number 4791304 Maintenance This section explains routine operator maintenance necessary to keep the Product in optimal condition Clean the Exterior To clean the Product wipe it with a cloth that is lightly dampened with water or mild detergen...

Page 32: ...es saturated When the entire normally blue desiccant material is pink the cartridge should be replaced See Display and Control Panel To change the desiccant cartridge see Figure 5 1 Turn the Product off 2 Turn the front panel cap turn counter clockwise and remove the cap 3 Loosen the rear seal of the desiccant by gently nudging the front of the desiccant in various directions up down left right 4 ...

Page 33: ...ar of the product to drain excess water from the Product See Display and Control Panel Warning To prevent possible electrical shock fire or personal injury do not overfill the reservoir Fuse Replacement The Product has mains fuses located in the mains input assembly The fuse rating label above the mains input assembly shows the correct replacement fuse ratings for each operating voltage Warning To...

Page 34: ...gure 6 1 Disconnect the mains power cord 2 With a standard screwdriver release the fuse holder door 3 Pull out the fuse holder 4 If necessary replace the fuse 5 Reinsert the fuse holder 6 Close the fuse holder door 3 2 1 ibw007 eps Figure 6 Replace the Mains Fuses ...

Page 35: ... Product has a fuse for the 12 V dc transducer power output located in the rear panel Replace only with a 1 A Fast Blow fuse F1A 250V 5x20mm To change the fuse see Figure 7 12VDC FUSE 1 AMP MAX GND DRAIN PLUG ibw024 eps Figure 7 Replace the 12 V Fuse ...

Page 36: ... uncertainty Michell Instruments S8000 Remote or RH Systems 473 with RP2 probe Readout 0 006 C 25 C thermistor uncertainty Fluke 1586A Super DAQ Precision Temperature Scanner with DAQ STAQ multiplexer Thermistor 0 01 C calibration uncertainty Fluke 5611T 8 probes are required Thermistor probe fixture Positions four of the thermistor probes in a square pattern at the rear of the working area and th...

Page 37: ...raw the sample from the chamber Experiment with the pump speed to draw enough gas for the chilled mirror to work but not so much that the Product performance is adversely affected A flow range of 0 1 liters to 0 5 liters per minute is usually adequate Use PTFE or stainless tubing for best results Caution When drawing gas out of the chamber in this method the dew point of the gas must be lower than...

Page 38: ... remote head into the door as shown in Figure 9 Ensure that the remote head sensor is inserted into the working volume of the chamber There is sufficient gas flow within the chamber such that an external pump is normally not needed in this situation ibw012 eps Figure 9 Chilled Mirror Probe Inserted Directly into the Door ...

Page 39: ... minute is usually adequate Ensure the chosen flow rate does not cause excessive chamber temperature non uniformity Use PTFE or stainless tubing for best results See Figure 10 Caution When drawing gas out of the chamber in this method the dew point of the gas must be lower than ambient temperature otherwise heated tubes must be used to prevent the formation of condensation in the tube ibw021 eps F...

Page 40: ...midity for that test point Table 11 Calibration Steps Step Temperature Set Point Humidity Set Point Dew Point Reading Temperature Reading Calculated RH Low T Tolerance High T Tolerance Low RH Tolerance High RH Tolerance 1 18 C 7 17 86 C 18 14 C 6 3 7 7 2 18 C 20 17 86 C 18 14 C 19 3 20 7 3 23 C 20 22 86 C 23 14 C 19 3 20 7 4 28 C 20 27 86 C 28 14 C 19 3 20 7 5 23 C 35 22 86 C 23 14 C 34 3 35 7 6 1...

Page 41: ... at the set point of 23 C 35 Calculate the difference from the reference as C C DUT_Actual_RH Ref_RH 7 Record the Product displayed Actual relative humidity at the set point of 23 C 80 Calculate the difference as D 8 Calculate RH_Error with RH_Average_Error C D 2 9 Calculate the new relative humidity offset with RH_Offset_New RH_Offset_As_Found RH_Average_Error 10 To update the temperature and rel...

Page 42: ...roduct off and check for condensation See Remove Condensation There is a leak in the system Check that the door is properly seated and tightened against the Product front panel Check that all ports are sealed properly Check Water indicator is lit with the water at the proper level Display has the message HI RH ERR Product setting is at a temperature and or humidity extreme relative to the prior se...

Page 43: ...hut off power Drain excess water using the drain outlet on the back of the Product until water level is between Max and Min with the Product in Standby or turned off Error Messages The product has Error Messages that show information during general use of the Product Error Messages are not necessarily related to the Error Codes that follow this section The latter indicates an internal failure that...

Page 44: ...rdware fault No Peltier Temp Internal over temperature probe fault fans will run continuously No HumGen Temp Internal temperature probe failure Peltier High Limit Internal cooling sensor is above limit HumGen Low Limit Ambient temperature of humidity generator is below operating temperature HumGen High Limit Humidity generator is above allowed limit Check water supply or ambient conditions EEPROM ...

Page 45: ...2 1 Mains Power Cord UK 769455 1 Mains Power Cord Swiss 769448 1 Mains Power Cord Australia 658641 1 Mains Power Cord South Africa 782771 1 Mains Power Cord Thailand 4362094 1 Mains Power Cord Denmark 2477031 1 Mains Power Cord Brazil 3841347 1 5128A Safety Information 4758293 1 FUSE 5 x 20 mm 4 Amp F4A 250V Fast Blow 3712591 2 FUSE 12 V dc output fuse F1A 250V fuse 3712540 1 Desiccant cartridge 4...

Page 46: ...MINIMUM MAXIMUM WORKING AREA CHAMBER COVER 1 00 inch 4 30 inches 4 88 inches FRONT RIGHT CORNER OF FLUKE 5128A FRONT PANEL FACE OF FLUKE 5128A Figure 11 Working Volume Template 42 ...

Reviews: