16
Probe Calibrate Screen
The probe calibrate screen shows the probe offsets for temperature and humidity
and is used in the Product-adjustment process. Use the probe calibrate screen to
load in a temperature offset and humidity offset to adjust the Product chamber
output. Probe offsets are entered as part of the routine calibration process.
An example of the calibration offset screen:
°
°
BIT Humidity Gen Screen
The BIT humidity gen screen contains parameters associated with the internal
humidity generator and is only used by trained service personnel. Push and hold
down
EXIT
to return to the main screen.
Error Code Screen
The error code screen is shown and errors that the Product detects are
presented. See
Error Codes
for a list of error codes. Push and hold
EXIT
to
return to the main screen.
Firmware Version Screen
Product firmware version details are shown. Push and hold
EXIT
to return to the
main screen.