Blank screen
The screen appears blank when switched on.
Ensure the power adapter is attached properly and the power is switched on.
This problem can also be caused by improper adjustment of the contrast. After the power
is switched on allow the thermo-hygrometer at least one minute to complete its self-test
and power on sequence. Then, try to adjust the contrast using the
A
D
buttons.
Error message at power up
The thermo-hygrometer reports an error during the power up self-test.
The display will remain inoperable during a power outage to preserve battery life.
Measurements will continue to be recorded as long as the battery has a charge.
On power up the thermo-hygrometer performs a series of self-tests to check the operation
of all its components and each of the sensors. The results of the self-tests are reported on
the screen. Normally, all tests report ‘OK’. Below is a table of error messages and
recommended actions to correct the error. Often these errors can be corrected by
completing the actions listed. A failure of one or more component causes error messages
to be reported.
Contact an Authorized Service Center for assistance.
Self-Test Error Messages
Table 8-1. Self-Test Error Messages
Error Message
Recommended Action
SYSTEMS: SERIAL PORT FAILURE
Cycle the power. Contact an Authorized Service
Center.
SYSTEMS: IR PORT FAILURE
Cycle the power. Contact an Authorized Service
Center.
SYSTEMS: DATA CARD FAILURE
Cycle the power. Contact an Authorized Service
Center.
BATTERY: FAILURE
Cycle the power. Contact an Authorized Service
Center.
BATTERY: NOT INSTALLED
Check that a battery is properly installed. Replace
the battery.
CLOCK: INVALID TIME
Set the time/date.
MEMORY DATA: EMPTY
Contact an Authorized Service Center.
MEMORY DATA: QUESTIONABLE
Cycle the power. Contact an Authorized Service
Center.
SENSOR <n>: NOT ATTACHED
Check that the sensors are properly inserted into
their sockets or attached with an undamaged
extension cable.
SENSOR <n>: LOCKED
Attach the correct sensor or, if a new sensor is to be
used, unlock the sensor using CHANNEL-SENSOR
LOCK.
Summary of Contents for 5020A
Page 6: ...5020A Users Manual iv...
Page 8: ...5020A Users Manual vi...
Page 12: ...5020A Users Manual 1 2...
Page 20: ...5020A Users Manual 2 2...
Page 24: ...5020A Users Manual 3 2...
Page 30: ...5020A Users Manual 3 8...
Page 32: ...5020A Users Manual 4 2...
Page 38: ...5020A Users Manual 5 2 Comm Setting 5 30 Serial 5 31 IR 5 32 Password 5 32 System Info 5 34...
Page 72: ...5020A Users Manual 6 2...
Page 124: ...5020A Users Manual 6 54...
Page 126: ...5020A Users Manual 7 2...
Page 130: ...5020A Users Manual 7 6...
Page 132: ...5020A Users Manual 8 2...
Page 138: ...5020A Users Manual 8 8...
Page 141: ...Index continued 3 W warnings 1 4 Z zone types 5 9...
Page 142: ...5020A Users Manual 4...