Insulation Multimeter
Electrical Specifications
41
Model 1587/1587 FC
Output Voltage
Display Range
Resolution
Test Current
Resistance Accuracy
±
(% of Rdg + Counts)
50 V
(0 % to +20 %)
0.01 to 6.00 M
Ω
0.01
M
Ω
1 mA @ 50 k
Ω
±
(3 % + 5 counts)
6.0 to 50.0 M
Ω
0.1
M
Ω
100 V
(0 % to +20 %)
0.01 to 6.00 M
Ω
0.01
M
Ω
1 mA @ 100 k
Ω
±
(3 % + 5 counts)
6.0 to 60.0 M
Ω
0.1
M
Ω
60 to 100 M
Ω
1
M
Ω
250 V
(0 % to +20 %)
0.1 to 60.0 M
Ω
0.1
M
Ω
1 mA @ 250 k
Ω
±
(1.5 % + 5 counts)
60 to 250 M
Ω
1
M
Ω
500 V
(0 % to +20 %)
0.1 to 60.0 M
Ω
0.1
M
Ω
1 mA @ 500 k
Ω
±
(1.5 % + 5 counts)
60 to 500 M
Ω
1
M
Ω
1000 V
(0 % to +20 %)
0.1 to 60.0 M
Ω
0.1
M
Ω
1 mA @ 1 M
Ω
±
(1.5 % + 5 counts)
60 to 600 M
Ω
1
M
Ω
0.6 to 2.0 G
Ω
100
M
Ω
±
(10 % + 3 counts)
Model 1577
Output Voltage
Display Range
Resolution
Test Current
Resistance Accuracy
±
(% of Rdg + Counts)
500 V
(0 % to +20 %)
0.1 to 60.0 M
Ω
0.1
M
Ω
1 mA @ 500 k
Ω
±
(2.0 % + 5 counts)
60 to 500 M
Ω
1
M
Ω
1000 V
(0 % to +20 %)
0.1 to 60.0 M
Ω
0.1
M
Ω
1 mA @ 1 M
Ω
±
(2.0 % + 5 counts)
60 to 600 M
Ω
1
M
Ω