1587 FC/1587/1577
Users Manual
20
Temperature (all 1587 Models)
The Meter measures the temperature of a type-K
thermocouple (included). Choose between degrees
Celsius (
°
C) or degrees Fahrenheit (
°
F).
1587 FC:
Press
to toggle between
°
C or
°
F.
1587/1587T:
Press
r
to toggle between
°
C or
°
F.
Caution
To prevent possible damage to the Meter or
other equipment, remember that while the
Meter is rated for -40
°
C to 537
°
C (-40
°
F to
998.0
°
F), the included K-type thermocouple is
rated for 260
°
C (500
°
F). For temperatures out
of that range, use a higher rated thermocouple.
Warning
To prevent risk of electric shock do not
connect thermocouple to electrically live
circuits.
80BK-A Type K
Thermocouple
Probe
Vent
or
Pipe
°C
°F
bav09f.eps
Figure 3. Measuring Temperature