110 PLUS and 115
Users Manual
20
General Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18
°
C to 28
°
C, with relative
humidity at 0 % to 90 %.
Extended specifications are available at
www.fluke.com
.
Maximum voltage between any terminal
and earth ground
........................................................... 600 V
W
Fuse for A input (115 only)
...................................... F 11 A, 1 000 V, Interrupt Rating: 17 kA
(Fluke PN 803293)
Display
........................................................................... Digital: 6 000 counts, updates 4/sec
Bar Graph: 33 segments, updates 32/sec
Temperature
.................................................................. Operating: -10 °C to + 50 °C
Storage: -40 °C to + 60 °C (limited by battery
specification)
Temperature Coefficient
............................................... 0.1 x (specified accuracy)/
°
C (<18
°
C or >28
°
C)
Operating Altitude
......................................................... 2 000 meters
Battery
........................................................................... 9 V Alkaline, NEDA 1604A / IEC 6LR61
Battery Life
.................................................................... Alkaline: 400 hours typical, without backlight
IP Rating (dust and water protection)
.......................... IP42 (non-operation)
Safety
............................................................................. IEC 61010-1: Pollution Degree 2
IEC 61010-2-033: CAT III 600 V