True-rms Multimeters

 

 Making 

Basic 

Measurements 

 

  

11 

Making Basic Measurements 

The figures on the following pages show how to make basic 

measurements.  

XW

 Warning 

To prevent possible electrical shock, fire, or 

personal injury: 

 

Disconnect power and discharge all high-

voltage capacitors before you measure 

resistance, continuity, capacitance, or a 

diode junction. 

 

Remove circuit power before you connect 

the Product in the circuit when you 

measure current. Connect the Product in 

series with the circuit. 

 

Do not use the HOLD function to measure 

unknown potentials. When HOLD is 

turned on, the display does not change 

when a different potential is measured.  

When connecting the test leads to the circuit or device, connect 

the common (

COM

) test lead before connecting the live lead; 

when removing the test leads, remove the live lead before 

removing the common test lead. 

Measuring Resistance 

 

edy04f.eps 

XW

 

Warning 

To avoid electric shock, injury, or damage to 

the Meter, disconnect circuit power and 

discharge all high-voltage capacitors before 

testing resistance, continuity, diodes, or 

capacitance. 

Summary of Contents for 110 PLUS

Page 1: ...er 2015 2015 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 110 PLUS and 115 True rms Multimet...

Page 2: ...ce center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNES...

Page 3: ...Alert 3 Safety Information 4 Display 7 Terminals 9 Rotary Switch Positions 10 Battery Saver Sleep Mode 11 MIN MAX AVG Recording Mode 11 Display HOLD 11 Backlight 11 Manual and Autoranging 11 Power Up...

Page 4: ...ng AC or DC Current 115 only 15 Measuring Current above 10 Amps 16 Measuring Capacitance 115 only 17 Measuring Frequency 115 only 17 Low Impedance Capacitance Measurements 115 only 18 Testing Diodes 1...

Page 5: ...o register your product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com usen support manuals Unsafe Voltage The hazardous voltage warning Y...

Page 6: ...esistant clothes to prevent shock and arc blast injury where hazardous live conductors are exposed Examine the case before you use the Product Look for cracks or missing plastic Carefully look at the...

Page 7: ...batteries when the low battery indicator shows to prevent incorrect measurements Remove the batteries if the Product is not used for an extended period of time or if stored in temperatures above 50 C...

Page 8: ...w voltage MAINS installation Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Categ...

Page 9: ...g Model The Meter function is set to Continuity 110 PLUS 115 The Meter function is set to Diode Test 115 Input is a negative value 110 PLUS 115 Y X Hazardous voltage Measured input voltage 30 V or vol...

Page 10: ...ing 110 PLUS 115 610000 mV Indicates the Meter s range selection 110 PLUS 115 Bar graph Analog display 110 PLUS 115 Auto Volts Auto Manual The Meter is in the Auto Volts function Autoranging The Meter...

Page 11: ...0 PLUS 115 Input terminal for measuring voltage continuity resistance capacitance frequency and testing diodes 110 PLUS 115 Error Messages Battery must be replaced before the Meter will operate Calibr...

Page 12: ...mV 110 PLUS 115 Ohms from 0 1 to 40 M 110 PLUS 115 Continuity beeper turns on at 20 and turns off at 250 110 PLUS 115 Diode Test Displays OL above 2 0 V 115 Farads from 1 nF to 9999 F 115 Hz button A...

Page 13: ...splayed Press to step through the low MIN average AVG and present readings To pause MIN MAX AVG recording without erasing stored values press is displayed To resume MIN MAX AVG recording press again T...

Page 14: ...he range does not change 3 To exit Manual Range press for at least 1 second or turn the rotary switch The Meter returns to Autorange and Auto is displayed Power Up Options To select a Power Up Option...

Page 15: ...current Connect the Product in series with the circuit Do not use the HOLD function to measure unknown potentials When HOLD is turned on the display does not change when a different potential is measu...

Page 16: ...eter s resistance e function Measuring AC and DC Voltage Volts DC Volts AC V edy03f eps Using Auto Volts Selection 110 PLUS only With the function switch in the position the Meter automatically select...

Page 17: ...Measuring AC or DC Current 115 only XW Warning To avoid personal injury or damage to the Meter Never attempt to make an in circuit current measurement when the open circuit potential to earth is 600 V...

Page 18: ...Current above 10 Amps The millivolt and voltage function of the Meter can be used with an optional mV A output Current Probe to measure currents that exceed the rating of the Meter Make sure the Mete...

Page 19: ...only edy05f eps Measuring Frequency 115 only XW Warning To avoid electrical shock disregard the bar graph for frequencies 1 kHz If the frequency of the measured signal is 1 kHz the bar graph and are u...

Page 20: ...nt Select progressively lower ranges using manual ranging for a stable reading Low Impedance Capacitance Measurements 115 only For making capacitance measurements on cables with ghost voltage hold whi...

Page 21: ...s disabled when measuring capacitance In frequency the bar graph and range annunciator indicates the underlying voltage or current up to 1 kHz The number of segments indicates the measured value and i...

Page 22: ...an repair the Product Use ONLY a fuse with the amperage voltage speed and interrupt rating specified To remove the battery door for battery replacement 1 Remove the test leads from the Meter 2 Remove...

Page 23: ...r and replace it with an 11 A 1000 V FAST fuse having a minimum interrupt rating of 17 000 A Use only Fluke PN 803293 6 To re assemble the Meter first attach the case bottom to the case top then insta...

Page 24: ...000 V Interrupt Rating 17 kA Fluke PN 803293 Display Digital 6 000 counts updates 4 sec Bar Graph 33 segments updates 32 sec Temperature Operating 10 C to 50 C Storage 40 C to 60 C limited by battery...

Page 25: ...k that supplies buildings used for domestic purposes There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Emis...

Page 26: ...110 PLUS 115 DC Volts 6 000 V 60 00 V 600 0 V 0 001 V 0 01 V 0 1 V 0 5 2 110 PLUS 115 DC 45 to 500 Hz 500 Hz to 1 kHz Auto V LoZ 1 True rms 600 0 V 0 1 V 2 0 3 4 0 3 110 PLUS 45 to 500 Hz 500 Hz to 1...

Page 27: ...shorts of 500 s or longer 110 PLUS 115 Ohms 600 0 6 000 k 60 00 k 600 0 k 6 000 M 40 00 M 0 1 0 001 k 0 01 k 0 1 k 0 001 M 0 01 M 0 9 2 0 9 1 0 9 1 0 9 1 0 9 1 5 2 110 PLUS 115 Diode test 2 000 V 0 0...

Page 28: ...ction Range Resolution Accuracy of Reading Counts Model AC Amps True rms 1 45 Hz to 500 Hz 6 000 A 10 00 A 3 20 A for 30 seconds max 10 minutes rest min 0 001 A 0 01 A 1 5 3 115 DC Amps 6 000 A 10 00...

Page 29: ...ange are not specified it is normal for this and other true rms meters to display non zero readings when the test leads are disconnected from a circuit or are shorted together For volts crest factor o...

Page 30: ...d Normal Mode Rejection Volts AC 5 M 100 pF 60 dB at dc 50 or 60 Hz Volts DC 10 M 100 pF 100 dB at dc 50 or 60 Hz 60 dB at 50 or 60 Hz Auto V LoZ 3 k 500 pF 60 dB at dc 50 or 60 Hz Open Circuit Test V...