Multimeters
Accuracy
Specifications
25
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C,
relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] +
[Number of Least Significant Digits])
Function Range Resolution
Accuracy
106 107
AC Volts (40
Hz to 500
Hz)
[1]
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3
1.0 % + 3
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 3
0.5 % + 3
AC Millivolts
600.0 mV
0.1 mV
3.0 % + 3
3.0 % + 3
Diode Test
[2]
2.000 V
0.001 V
10 %
[1] All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
[2] Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
106 & 107
Users Manual
26
Function Range Resolution
Accuracy
106 107
Resistance
(Ohms)
400.0
Ω
4.000 k
Ω
40.00 k
Ω
400.0 k
Ω
4.000 M
Ω
40.00 M
Ω
0.1
Ω
0.001 k
Ω
0.01 k
Ω
0.1 k
Ω
0.001 M
Ω
0.01 M
Ω
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3
Capacitance
[1]
50.00 nF
500.0 nF
5.000
μ
F
50.00
μ
F
500.0
μ
F
1000
μ
F
0.01 nF
0.1 nF
0.001
μ
F
0.01
μ
F
0.1
μ
F
1
μ
F
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5
Frequency
[2]
Hz
(10 Hz – 100
kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
NA
0.1 % + 3
Duty Cycle
[2]
1 % to 99 %
0.1 %
NA
1 % typical
[3]
[1] Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the
50 nF range).
[2] All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
[3] Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %