background image

1000FLT 

Users Manual 

16 

Specifications 

Temperature 

Operating .......................... -10 

°

C to +50 

°

C (+14 

°

F to +122 

°

F) 

Storage ............................. -40 

°

C to +60 

°

C (-40 

°

F to +140 

°

F) 

Relative Humidity

 ................ 85  % 

Operating Altitude

 .............. <2000 m 

Battery 

Type .................................. 4 x AA Alkaline IEC LR6 
Life .................................... 40  hr 

NCV (VoltAlert

)

 ................. 85 V ac to 400 V ac 

45 Hz to 67 Hz 

10 cm (4 in) distance 

Pin Continuity Test

 ............. <1  k

Ω

 

Ballast Test

 .......................... 20  kHz

 

Ballast Type 

Electronic .......................... 20  kHz 
Magnetic ........................... 50 Hz to 60 Hz 

3 m (10 ft) distance 

Drop

 ..................................... 2 meter 

Size (H x W x L)

 ................... 21.5 cm x 3.0 cm x 6.5 cm  

(8.5 in x 1.2 in x 2.6 in) 

Weight

 .................................. 0.37 kg (0.80 lb) 

Warranty

 

Main  Body ......................... 3  years 
Rod  Extender .................... 1  year 

 

Safety

 ................................... IEC 61010-1, Pollution Degree 2

 

Electromagnetic Compatibility (EMC) 

EU ..................................... IEC 61326-1 (Portable) CISPR 11, Group 2, 

Class A.  

 

Group 2 equipment contains all ISM RF 
equipment in which radio-frequency energy 
in the frequency range 9 kHz to 400 GHz is 
intentionally generated and used or only 
used, in the form of electromagnetic 
radiation, inductive and/or capacitive 
coupling, for the treatment of material or 
inspection/analysis purposes. 

 

Class A equipment is suitable for use in all 
establishments other than domestic and 
those directly connected to a low voltage 
power supply network which supplies 
buildings used for domestic purposes. 
Caution, there may be potential difficulties in 
ensuring electromagnetic compatibility in 
other environments, due to conducted and 
radiated disturbances.

 

US (FCC) .......................... 47 CFR 15 subpart B, this product is 

considered an exempt device per clause 
15.103 

Applies to use in 
Korea only (KCC) .............. Class A Equipment (Industrial Broadcasting 

& Communication Equipment).  

 

This product meets requirements for 
industrial (Class A) electromagnetic wave 
equipment and the seller or user should 
take notice of it. This equipment is intended 
for use in business environments and not to 
be used in homes. 

Summary of Contents for 1000FLT

Page 1: ...14 2014 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1000FLT Fluorescent Light Tester Users...

Page 2: ...ing the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem T...

Page 3: ...efore You Start 3 Power On and Off 3 Battery 3 Beeper On and Off 4 Rod Extender 4 Controls 5 Lamp Test 6 Ballast Test 7 Non Contact Voltage Test VoltAlert 8 Pin Test 10 Ballast Type Test 11 Product Ca...

Page 4: ...1000FLT Users Manual ii...

Page 5: ...ii List of Tables Table Title Page 1 Symbols 3 2 1000FLT Features and Controls 5 3 Lamp Test 6 4 Ballast Test 7 5 Non Contact Voltage Test 9 6 Pin Test 10 7 Ballast Type Test 12 8 Replaceable Parts 13...

Page 6: ...1000FLT Users Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 Rod Extender 4 2 Optical Sensor for Ballast Type Test 11 3 Battery Replacement 14 4 Rod Extender Replacement 15...

Page 8: ...1000FLT Users Manual vi...

Page 9: ...e s website at www fluke com To register your Product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com usen support manuals Safety Informati...

Page 10: ...ed users that prevent an effective ground Receptacles in recessed sockets and or differences in socket design Condition of the Product and batteries Use the Product only as specified or the protection...

Page 11: ...toring and Control Instrumentation product Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Before You Start The purchase of your Product incl...

Page 12: ...Warning To prevent possible electrical shock or personal injury Do not touch the rod extender while performing any test Do not allow the rod extender to contact live voltage Completely collapse the ro...

Page 13: ...t results Table 2 shows the location and describes each function Table 2 1000FLT Features and Controls huz01 eps Item Description Power On Off Test Function Buttons Rod Extender Saddle Optical Sensor...

Page 14: ...on and the LED test indicator blinks green Product is ready to test 3 Continue to push and hold for the entire test duration 4 Place the saddle end of the Product on the lamp and continue to hold If...

Page 15: ...1 second The green light around the button perimeter turns on and the LED test indicator blinks green 3 Pull out the rod extender and scan 10 cm 4 in at both ends of the fluorescent bulb A solid green...

Page 16: ...er during the VoltAlert test Do not allow the rod extender to contact live voltage When using VoltAlert if the test indicator does not glow or the beeper does not sound hazardous voltage could still b...

Page 17: ...n of the test object A solid red LED and a continuous beep if beeper is on indicates that line voltage has been found The indicator continues to blink red if no line voltage is found Note The Product...

Page 18: ...hat is lit To test see Table 6 1 Turn on the Product 2 Insert the lamp pins into the slots on the Product The green LED test indicator shows the filament is good You will hear a continuous beep if it...

Page 19: ...ronic see Table 2 for the sensor location Make sure that the entire sensor is in line with the fluorescent light fixture as you test See Figure 2 The fluorescent light fixture must be turned on for th...

Page 20: ...of any part of the lit fluorescent lamp The green LED test indicator shows the test is complete See indicator for ballast type No color on the LED test indicator shows a test error Table 7 Ballast Ty...

Page 21: ...ts Have an approved technician repair the Product Table 8 is a list of replaceable parts available from Fluke Table 8 Replaceable Parts Description Fluke Part Number Holster 4532652 Rod Extender Saddl...

Page 22: ...eries when the light is blinking green and the Product automatically turns off To replace see Figure 3 1 Loosen the capture screw on the battery door and remove 2 Remove the spent batteries and dispos...

Page 23: ...door 2 Remove the batteries 3 Remove the screw from the base of the rod extender 4 Pull out the faulty rod extender from the body of the Product 5 Slide the replacement rod extender into place 6 Repl...

Page 24: ...range 9 kHz to 400 GHz is intentionally generated and used or only used in the form of electromagnetic radiation inductive and or capacitive coupling for the treatment of material or inspection analys...

Page 25: ...014 2014 Fluke Corporation All rights reserved 1000FLT Fluorescent Light Tester Quick Reference Guide Warning To prevent possible electrical shock fire or personal injury read the 1000FLT Safety Sheet...

Page 26: ...1 2 3 NCV VoltAlert Ballast Type Test 1 sec 2 3 180 1 sec 1 2 3 1 Rod Extender Replacement...

Reviews: