HF110AEN_v0501_04
Page 44
VAR DESCRIPTION
BYTES
VALUE
REMARKS
DISPLAY
64
(40h)
display function
1
0=total
1=flowrate
68
(44h)
set flowrate monitor 1
0=operator level
1=SETUP level
POWERMANAGEMENT
80
(50h)
LCD update time
1
0=fast
1=1sec
2=3sec
3=15sec
4=30sec
5=off
81
(51h)
power-mode battery 1
0=operational
1=shelf
FLOWMETER
98
(62h)
formula 1
0=linear
1=square root
99
(63h)
filter 1
0….99
100
(64h)
cut-off
2
0….999
steps of 0.1%
102
(66h)
calibration low
(4mA)
1 0=default
1=calibrate
2=cal set
103
(67h)
calibration high
(20mA)
1 0=default
1=calibrate
2=cal set
ANALOG OUTPUT
112
(70h)
analog output
1
0=disable
1=enable
113
(71h)
minimum rate
3
0..9999999
unit, time, decimals acc. var48-50
116
(74h)
maximum rate
3
0..9999999
unit, time, decimals acc. var48-50
119
(77h)
cut off percentage
1
0..99
steps of 0.1%
120
(78h)
tune minimum rate
2
0..9999
122
(7Ah)
tune maximum rate 2
0..9999
99
(63h)
filter 1
0….99
PULSE OUTPUT
128
(80h)
impulse width
1
0=off
1=short
2=long
129
(81h)
pulse per X quantity 3
1..9999999
unit, decimals acc. var32 -33
OTHERS
168
(A8h)
pass code
2
xxxx
read only!
170
AAh
tagnumber
3
0..9999999
Other vars: see standard table