![Fluidwell F014-P Manual Download Page 31](http://html1.mh-extra.com/html/fluidwell/f014-p/f014-p_manual_2308987031.webp)
Page 30
FW-F014-P-M_v0402_05_EN.docx
Label information pulse input type - F0..P-XI (inside and outside the enclosure)
Fig. 28: Label information - Intrinsically Safe application.
5.2
TERMINAL CONNECTORS INTRINSICALLY SAFE APPLICATIONS
Terminal connectors F014-P-(PC / PD / PX)-XI-(ZB):
Fig. 29: Overview terminal connectors XI - Intrinsically Safe application.
•
Type PC:
offers - additional to type PX - an internal Intrinsically Safe lithium battery. This ATEX
certified battery (FW-LiBATT-xxx) may be changed in hazardous area.
•
Type PD:
offers with terminal 6 - additional to type PX - a real sensor supply of 8.2V DC to power a
Namur sensor for example.
•
Type PX
:
as standard, all intrinsically product are supplied with terminal 4 and 5 to power the
product externally.
Made in Veghel, The Netherlands by Fluidwell bv
IP67
Tamb
-40°C ... +70°C
-40°F ... +158°F
MODEL: F0xx-P-XI
Signal input: pulse, coil, namur
Intrinsically Safe - Possible static hazard - Do not rub
KEMA 05ATEX1168 X
IECEx KEM 08.0006X
II 1 G Ex ia IIC T4
II 1 D Ex iaD 20 IP 65 / 67 T 100°C
Ga Ex ia IIC T4
Ex iaD 20 IP 65 / 67 T 100°C
Intrinsically Safe for Class I/II/III, Div. 1, Grps A,B,C,D,E,F,G,
Temp. class T4 – Class I, Zone 0, AEx ia IIC T4 - For installation
and entity evaluation refer to control drawing no. FWCD-0003.
CSA Certificate nr: CSA.08.2059461 – FM Project ID: 3033306
Made in Veghel, The Netherlands by Fluidwell bv
MODEL: F0xx-P-XI
Signal input: pulse, coil, namur
Intrinsically Safe - Possible static hazard - Do not rub
KEMA 05ATEX1168 X
Tamb
IP67
II 1 G Ex ia IIC T4
II 1 D Ex iaD 20 IP 65 / 67 T 100°C
IECEx KEM 08.0006X
Ga Ex ia IIC T4
Ex iaD 20 IP 65 / 67 T 100°C
Battery powered applications:
Replace with certified
FW-LiBAT-00x battery only!
Remark: consult the manual.
Intrinsically Safe for Class I/II/III, Div. 1, Grps A,B,C,D,E,F,G,
Temp. class T4 – Class I, Zone 0, AEx ia IIC T4 - For installation
and entity evaluation refer to control drawing no. FWCD-0003.
CSA Certificate nr: CSA.08.2059461 – FM Project ID: 3033306
-40°C ... +70°C
-40°F ... +158°F