CUT-OFF TIME
2.7
With this setting, a minimum flow requirement threshold is determined, if
during this time less than XXX-pulses (SETUP 2.6) are generated, the
flowrate will be displayed as zero.
The cut-off time has to be entered in seconds - maximum time is 999.9
seconds (about 15 minutes).
3 - DISPLAY
FUNCTION
3.1
The large 12mm (
0.47”)
digits can be set to display total, flowrate or
accumulated total.
When "total" is selected, total is displayed with the large 12mm (
0.47”)
digits and flowrate is displayed with the 7 mm (
0.28”)
digits
simultaneously.
When "flowrate" is selected, only flowrate will be displayed with the large
12mm (
0.47”)
digits together with its measuring unit while total will be
displayed after pressing SELECT.
When "accumulated total" is selected, total is displayed with the large
12mm (
0.47”)
digits and accumulated total is displayed with the 7 mm
(
0.28”)
digits simultaneously.
When battery powered, the user can expect reliable measurement over a long period of time. The
E112-P has several smart power management functions to extend the (optional) battery life time
significantly next to permanently disabling the infra-red keys. The following functions can be set:
LCD NEW
3.2
The calculation of the display-information influences the power
consumption significantly. When the application does not require a fast
display refresh rate, it is
strongly advised
to select a slow refresh rate.
Please understand that NO information will be lost; every pulse will be
counted and the output signals will be generated in the normal way.
The following refresh rates can be selected:
fast - 1 sec - 3 sec - 15 sec - 30 sec - off.
(read 3 sec as a refreshment every 3 seconds, 30 sec as a refreshment
every 30 seconds)
When “off” is selected the display will be automatically switched off after
30 seconds since the last action. The display will return again after
touching one of the keys.
BACKLIGHT
3.3
The unit is provided with a backlight. The backlight brightness can be
adjusted from off (0) to a 100% in steps of 20%.
The following can be selected:
0
%
– 20
%
– 40
%
– 60
%
– 80
%
– 100
%
For battery powered untis the brightness will not change.
Note:
When the unit is only loop powered it cannot make use of the
backlight. In that case an external supply is required.
BARGRAPH
3.4
The unit has a speedometer display which
offers a quick impression of the
actual flow rate.
This graph on the display can be switched on or off.
The following can be selected:
enable – disable
RATESPAN
3.5
With Ratespan the range of the speedometer can be set. The display
shows 0 – 100 with 20 blocks so each block is 5% of the 100% range set
here. The following can be selected:
0000.000 - 9999999
The number of decimals displayed depends upon SETUP 2.3.