PROMETRA® II PROGRAMMABLE PUMP
Page 26 of 38
Demand Bolus Regimen
Pre-Programmed Pump Settings
When the Prometra Programmer inquires the Prometra II Pump for the first time, the pump status
screens display the pre-programmed pump settings. The clinician can change these presets using the
programmer.
Parameter
Description
Data Preset
Patient
Name or code
No
Pump Model
Model of Pump
Yes, Prometra II
Pump SN
Serial Number of Pump
(e.g., 36DA4A77)
Yes, pump specific
Pump Ver.
Current control software version
(e.g. 1.02)
Yes, pump specific
Drug
Drug contained in pump
No, specified by user
Conc
Concentration of drug in pump
Yes, preset to 1.000 mg/mL
Accum
Accumulator Volume Constant (e.g.
2.010 μL)
Yes, pump specific
Reservoir Volume Current estimated volume
contained in reservoir
Yes, 00.0 mL
Low Res. Alarm
Alarm to indicate reservoir volume
is low
Yes, to OFF
Low Res. Volume Setting to actuate Low Reservoir
Alarm
Yes, 2.0 mL
Battery
Pump battery charge status
No, reports condition, e.g.
OK or Low
Flow Mode
Constant Flow, Multiple Rates,
Periodic Flow or Demand Bolus
Yes, to Constant Flow
Next Refill
Date: month/day/year (mo/da/yr)
Calculated by programmer,
appears after refill is programmed
No
Daily Dose
Programmed daily dose
(e.g. 2.005 mg)
Yes, to 0.000 mg