PROMETRA® II PROGRAMMABLE PUMP
Page 15 of 38
Emergency Procedure PRE-MRI Steps for Prometra II Pump
1.
Pump Inquiry
a.
Inquire the pump with the programmer to verify pump model, the pump is operational
and without errors.
b.
Verify that more than 1mL of drug is present in the Drug Reservoir.
c.
Print inquiry page.
2.
Pump Programming
a.
Set the flow mode to a constant flow rate of 0.0 mg/day.
b.
Re-inquire the pump and print inquiry page to confirm a constant flow rate of 0.0 mg/day.
Emergency Procedure POST-MRI Steps for Prometra II Pump
1.
Confirm Pump Operational Status –
a.
Inquire the pump with the programmer to verify pump operation and settings.
b.
Confirm that settings are unchanged from the Pre-MRI settings, e.g., flow rate must be 0.0
mg/day.
c.
If the programmer displays any pump errors, proceed to Step 2 “Clear Pump Errors”.
d.
If no pump errors are displayed, proceed to Step 3 “FAV Reset Procedure”.
2.
Clear Pump Errors
a.
If pump errors are displayed from the Inquiry performed in Step 1, perform an Emergency Pump
Stop using the programmer, and contact Flowonix Technical Solutions for assistance 855-356-9665.
b.
If pump errors are cleared, proceed to Step 3.
3.
FAV Reset Procedure
a.
Remove drug from Drug Reservoir by aspirating through the Refill Port.
b.
To aspirate, attach the 22G non-coring needle to a syringe barrel (available in Refill Kit).
c.
Advance needle through the center Refill Port Septum until needle tip resides completely inside the
Drug Reservoir.
WARNING: IF PUMP STATUS CANNOT BE PROPERLY CONFIRMED, DO NOT
PROCEED SINCE
THE PUMP MAY NOT BE OPERATING PROPERLY, PLEASE CONTACT FLOWONIX TECHNICAL
SOLUTIONS FOR ASSISTANCE AT: 855-356-9665.
WARNING: IF PUMP STATUS CANNOT BE PROPERLY CONFIRMED, DO NOT
PROCEED SINCE
THE PUMP MAY NOT BE OPERATING PROPERLY, PLEASE CONTACT FLOWONIX TECHNICAL
SOLUTIONS FOR ASSISTANCE AT: 855-356-9665.