68
−
the level of electromagnetic interference created by surrounding equipment, should
not exceed a permissible level for the meter.
Before hydraulic zero setting it is necessary to run the water meter in waterworks on the
maximal water flow rate for half an hour, then to cut off a valve after flow measurement sec-
tion, to cut off a valve before flow measurement section. Hydraulic zero setting can be con-
ducted in a few minutes after water fluctuations in the tapped section will stop.
If value of the first group of digits, displayed on the meter indicator and registered by
the meter as water zero velocity noticeably exceeds 500 or number of measurement cycles,
then some obstacles are possible on tapped FS:
−
air bubbles in FS;
−
water leak through valves;
−
significant level of external electromagnetic noise.
Summary of Contents for SVTU-11B
Page 1: ...Water meter SVTU 11B Operating manual SMP 407251 010 Part 1 September 2010...
Page 2: ......
Page 29: ...29 Figure 9 7 Installation of R D S of type 2 3 and 4 without thermal pocket...
Page 30: ...30 Figure 9 8 Installation of R D S of type 2 3 and 4 with thermal pocket...
Page 37: ...37 Figure 11 1...
Page 40: ...40 Figure 121 1...
Page 49: ...49 Appendix C Overall and connecting dimensions of calculator 17...
Page 51: ...51...
Page 52: ...52 Indication of main parameters mode...
Page 53: ...53 Check mode...
Page 54: ...54 Setup mode...
Page 55: ...55 Setup Set zero mode Setup PS parameters mode...
Page 56: ...56 Setup Pressures mode...
Page 61: ...61 Schematic of device cable...
Page 62: ...62 Pinout of connectors...
Page 64: ...64 Appendix L Overall and connecting dimensions for flow measurement sections FS...
Page 65: ...65 Overall and connecting dimensions for FS 32...