*M12
Input pipe wall thickness.
*M13
Input inner diameter
*M14
Select pipe material
M15
Input sound velocity of the pipe material
M16
Select lining material
M17
Input sound velocity of the lining material
M18
Input the lining thickness
M19
Input the absolute roughness of pipe inner wall
*M20
Select the liquid type
M21
Input sound velocity of the liquid
M22
Input viscosity of the liquid
M09
Display today’s total NET flow
M10
Input outer perimeter
*M11
Input outer diameter. Available range is 0 to 18000mm
*M23
Select the transducer type. Over 20 types can be selected
*M24
Selec the mounting method of transducer
*M25
Display the transducer mounting distance
*M26
0 Use RAM Setting :
A switch for the parameters in flash memory will be loaded
when power is turned on. The default option is that the parameters will be load
If this switch is not turned on, the system will try to use the parameters are ok,
otherwise the system will load the parameters in flash memory.
1 Solidity setting:
Function to store the current parameters into the flash memory
, so that these parameters will be solidified and will be loaded as the default
parameters every time when power is turned on
.
M27
To save the installation point parameter.
M28
Maintain the last good value when poor signal condition occurs. YES is the
default setup.
M29
Setup a signal strength as empty pipe. For example input 65, it means the pipe is
treated as
empty pipe when the signal is less than 65, and the flow rate display is 0.
M30
Select metric or british system unit.
M31
Select flow rate unit.
M32
Select totalizer flow unit.
M33
Select totalizer multiple factor.
The multiplying factor ranges from 0.001 to 10000. Factory default is 1
M34
Turn on or turn o ffthe NET totalizer
M35
Turn on or turn o ffthe POS (positive) totalizer
M36
Turn on or turn off
the NEG(negative) totalizer
M37
Restore factory settings. Totalizer reset.
M38
Manual totalizer used for easier calibration. Press a key to start and press a key to stop
the
manual totalizer.
M39
Language selection. there are Italian or Turkish
*M40
Damping factor. The damping factor ranges from 0 to 999 seconds.
0 means there is no damping. Factory default is 10 seconds
*M41
Cut-o fflow flow rate (or zero flow rate) to avoid invalid accumulation.