>> Operation steps
·
Tighten the locknut into ball valve, adjust the insertion depth scale.
·
Open ball valve, completely push in the upstream transducer rod. Adjust the
transmit direction parallel with pipe axis, and point to the
installation point of
downstream transducer.
Lock it after adjustment.
·
Install downstream transducer in the same way. Adjust the transmit direction to
get the best signal strength and watching Menu91, if the value is between 97% ~
103%, the installation is
correct. If not, need to re - adjust the insertion depth and transmit direction until
meet the requirement.
7. Check installation
Please see details in Chapter 7.5
7.4
In
-
line type transducer installation
After choosing the installation point, install the transducer in the pipeline with companion
flanges.
Then connect the transducer to converter with special signal cable. Installation is complete.
1) Installation method
>> Transmit directon
There is a indicatng arrow on the transducer junccon box, the arrow
directon on two
transducers should be opposite and parallel to the pipe
axis.