>>Pressure
The maximum press ure for standard insertion and inline transducer is
1.6MPa
Out of this range need customized.
>>EMI (electromagnetic interference)
The ultrasonic flow meter, transducer and signal cable can be easily interfered by interference
sources such as frequency changer, radio station, microwave station, GSM base station and
high-tension cable. Please try to avoid these interference sources when choosing installation
points.
The shield layer of flow meter, transducer and signal cable should be connected to earth.
Better to use isolated power supply. Do not use the same power supply with the frequency
converter.
>>Instrument well
When measuring underground pipes or need to protect the measuring points, an instrument
well is required. To ensure the enough installation space, the sizes of instrument well should meet
the following requirements.
7.2
Clamp on transducer Installation
Before installation, please verify the parameters of pipeline and liquid. To ensure the
installation accuracy.
1)
Installation procedure
Select an installation method → Input the measuring parameters → Clean pipe surface →
Install transducers → Check the installation
2)
Select an installation method
There are two different methods for clamp on transducers: V method and Z method.
>> V method
V method should be priority selected for pipe sizes DN15 -
DN200. Let the pair of
transducers horizontal alignment, the central line in parallel with the pipeline axis.