- 10
9. Maintenance
9.1 Check the running status of the machine before starting it, ensure that the power
supply of the device is reliably grounded. Normal operation can be put into use. The
external wiring of the equipment must be operated in the power-off state to prevent
burning the equipment.
9.2 Check to see if the tubing is damaged or the equipment is leaking and ensure that
the liquid in the tubing is discharged to the container or discharge pipe. Wipe any
leaked liquid around the pump quickly and correct any possible faults promptly. Keep
the rollers of the pump head clean and dry, otherwise it can hasten tubing wear (it can
also damage the tubing causing solid substances in the liquid), reduce the useful life
of tubing, and lead to quicker damage and wear to the rollers. If the pump head gets
water on it accidently, use a soft, absorbent cloth to wipe it dry to prevent damage to
the pump head. If liquid splashes on the pump, please turn off the power supply and
unplug the power wiring. Check to see if there is liquid in the inner part of the
equipment, if so please contact the manufacturer.
9.3 The factors affecting the flow are as follows:
1)
The inner diameter and wall thickness of tubing may have some deviations due
to its tolerance and different bathes.
2)
Different liquid properties
3)
It will decrease flow rate when input port and output port change incrementally
or there are suction and lift.
4)
The speed of DC/AC motor will change according to load, temperature and
humidity.
This may influence product flow accuracy, please leave a margin when choose tubing.
9.4 This product has no waterproof measures. Please take protective measures when
using in a wet environment. When the product will not be in use for a long time, please
clean it and keep it in a dry and ventilated environment.
9.5 Pump head is not resistant to organic solvent (Except for special indication) and
strong corrosive liquids, please be attentive when using either near it.