background image

FCS  KA  (SS)

- 1 - 

1. Product Introduction

FCS KA (SS), with low noise and stable operation, is driven by an integrated stepper
motor with a compact structure. The product adopts OLED blue screen display and
has an external control interface, which can realize remote control. It can communicate 
with a computer through standard Modbus communication protocol (RTU mode).

2. Product Appearance

A
B

C

D

F

E

G

H

A—— Operation panel 
B—— Drive 
C—— Pump head 
D—— Fan interface 
E—— Power supply switch 
F—— USB interface 
G——External control interface 
H——Power supply socket 

Summary of Contents for FCS KA

Page 1: ...FCS KA SS Compact Peristaltic Pump Operation Manual ...

Page 2: ...f the wire If following situations happen please turn off the power supply and disconnect the plug Hold the plug instead of the wire 1 Fluid splash on the pump 2 You think the pump needs to be maintained or repaired The user s power socket must have a ground wire with a reliable grounding Note The foot pedal switch and other external control plugs must be connected and disconnected in the power of...

Page 3: ...duct Appearance 1 3 Operation Panel 2 4 External Control Instruction 3 5 Communication Function Instruction 7 6 Technical Specification 8 7 Main Function and Feature 9 8 Dimension Drawing 9 9 Maintenance 10 10 Warranty After Sales Service 12 ...

Page 4: ... OLED blue screen display and has an external control interface which can realize remote control It can communicate with a computer through standard Modbus communication protocol RTU mode 2 Product Appearance A B C D F E G H A Operation panel B Drive C Pump head D Fan interface E Power supply switch F USB interface G External control interface H Power supply socket ...

Page 5: ...ys Direction This button is used to change the motor running direction Press this button once and the running direction of pump changes once External control button This button is used to change the method of analog speed control Press this button once and that will cycle the display Empty 5V 10V mA Empty Respectively corresponding to Turning off analog speed regulation 0 5V 0 10V voltage analog s...

Page 6: ...b the lower right corner of the OLED flashes and the value is RS485 communication address 1 32 At this time you can rotate the digital knob to change the value After setting press the digital knob to save the data and the screen no longer flashes 4 External Control Instruction FCS KA SS series has a 15 pin interface for the external control interface as the below figure shows That includes the fun...

Page 7: ...hort circuit break pulse mode is start short circuit break again is stop Pin2 to Pin14 short circuit break pulse mode is direction Level mode Pin3 is external control level mode switching Short circuited Pin3 and Pin14 to switch to external control level mode Pin1 to Pin14 short circuit is start breaking them is stop Pin2 to Pin14 short circuited motor is counterclockwise breaking them motor is cl...

Page 8: ...og speed regulation signal of the pump to 0 5V voltage signal by external control button The analog signal is displayed 5V in the screen Voltage 0 5V speed regulation Pin6 connects to the positive pole of voltage analog 0 5V signal Pin9 to the connects negative pole Change the voltage value of the analog signal and the speed changes linearly 0V corresponds to speed 0rpm 5V corresponds to maximum s...

Page 9: ...peed Note DO NOT connect the 0 5V 0 10V signal with 4 20mA input terminal Wrong connections may cause damage to the device The external control plugs must be connected and disconnected in the power off status to prevent the external control interface from being burned 4 Output signal 1 Connect Power VDD with the positive pole of external connection power supply 5V 2 Connect Power GND with the nega...

Page 10: ...is USB B Connect the pump with a computer through the USB interface Type A connects computer USB interface Type B connects KA SS USB interface Type Type A Type B Plug male USB2 0 4 3 2 1 Type A 1 2 4 3 Type B Connect with computer Connect with external control USB interface RS485 communication Connect A pin5 from external control interface as shown in Picture to T R of RS485 module and connect B p...

Page 11: ...xternal control Start stop Switch signal Speed control 0 5V 0 10V 4 20mA for option Speed resolution 0 1rpm Communication RS485 Control Digital knob pure imported keypad Temperature 0 40 Motor 57 closed loop stepper motor Relative humidity 80 Display OLED Output interface Output motor working status Power consumption 75W IP rate IP31 FCS KA SS ...

Page 12: ... at the same time RS485 address cannot be repetitive The circuit has excellent performance good heat dissipation conditions low working noise stable operation and has power off memory function RS485 standard Modbus protocol that is more convenient to connect with PLC 304 stainless steel shell anti corrosion no rust conforms to GMP requirements OLED displays motor speed and working status 8 Dimensi...

Page 13: ...age to the pump head If liquid splashes on the pump please turn off the power supply and unplug the power wiring Check to see if there is liquid in the inner part of the equipment if so please contact the manufacturer 9 3 The factors affecting the flow are as follows 1 The inner diameter and wall thickness of tubing may have some deviations due to its tolerance and different bathes 2 Different liq...

Page 14: ...ating current 9 8 This product does not have special certification such as medical certification When it needs to be used in special fields such as medical and military please self certify Load changes will also cause changes in operating current Therefore please conduct margin design for power supply power 9 9 The company shall not bear the direct and indirect losses caused by the malfunction or ...

Page 15: ...od after confirmation by our technical department we will provide spare parts free of charge Customers will need to bear the shipping cost Exceptions The warranty shall not apply to repairs or service necessitated by normal wear and tear or for lack of reasonable and proper maintenance All tubing and pumping accessories as consumable items are excluded Electrical surge as a cause of failure is exc...

Reviews: