4001V Peristaltic Pump O&M Manual
5.3.5 Interfacing
A dual 8 Amp DC DPDT relay is provided to signal or control an external device during an alarm condition.
Connect one set of the Tubing Rupture Detector NO relay contacts to pins 7 & 8 of the K4 SERIES pump
controller.
Program the K4 SERIES Pump Controller step 7.06 to 0008 for external leak fault to stop the drive in case of a
tubing rupture.
DUAL PUMP SYSTEMS
You may optionally connect the second set of relay contacts to pins 16 & 4 of the backup pump K4 SERIES
Pump Controller to make that pump take over when the first pump stops.
5.3.6 Calibration
Turn the P1 sensitivity adjusting screw counter clockwise several turns. Apply a wet rag to the LIQUID
SENSOR electrodes. Adjust the sensitivity clockwise slowly until the alarm trips.