Introduction      

 

1.2    Overview 

The Insertion Meter provides a cost effective and simple means of measuring the flow of a 
wide range of low viscosity liquids. Installation is quick and inexpensive for pipe diameters 
ranging from 40mm to 900mm (1.5-36") and up to 2500mm (100") nominal bore for the Hot 
tap capable model DP525.  
 
The Insertion Meter has a linear measuring range of  0.3~10.0 metres/sec. (1~33 ft/sec.). 
Minimum detectable flow velocity is 0.15 m/sec. (0.5 ft/sec.). When used in conjunction with 
the RT12 flow rate totaliser NLC feature the linear flow range is extended down to 0.15 
m/sec. (0.5 ft/sec.) with an improved linearity. 
 
The Insertion Meter is constructed from 316 L (1.4404) stainless steel enabling use in many 
applications for metering water and low viscosity chemicals. 
  
Two independent pulse outputs are standard & can directly input to a wide range of ancillary 
instruments, PLC’s and computers. Both pulse outputs have a high level of immunity to 
electrical interference. Options include a reed switch. 

 
1.3  Operating principle 

Flow passes through a pipe causing the rotor to spin. Magnets installed in the rotor pass by 
pulse sensors within the transducer body & inturn this produces frequency outputs 
proportional to flow rate. 

 

1.4  Specifications       

( subject to change without notice ) 

Model  

490                             525 

Suits pipe sizes 

40mm - 900mm                           50mm - 2500mm 

( 1.5 

– 36 inches )                        ( 2 – 100 inches ) 

Flow range 

0.25 - 6300 litres/sec.                0.4 - 49000 litres/sec. 

( 4-99600 USGPM )                      ( 6-780000 USGPM ) 

Process connections 

     

1.5" or 2” NPT or BSPT                   2" NPT or BSPT  

Velocity range 

0.3 - 10 metres/sec.   ( 1 - 33 feet/sec.) 

Linearity 

typically 

 1.5% O.R. (1-10m/s)  

 1.5% F.S. (0.3-1m/s) 

Repeatability 

typically 

 0.5% 

Pressure (max) 

80 Bar   ( 1200PSI ) 

Temperature range 

-40

C to 100

C (-40

F to 212

F ) standard 

Other temperature ranges available 

Body material 

316L stainless steel  (1.4404) 

Rotor materials 

Carbon Fibre filled PVDF 

(standard) 

or optional PEEK rotor with PEEK-HPV bearing  

O-Ring material 

VITON -  options available 

ELECTRICAL 

 

(a)

 Square wave ( Hall Effect ) 

5-24vdc, 3wire NPN open collector (20mA max. current sink) 

(b)

 Reed Switch  (to 100°C) 

30vdc max. x  20mA max.  (output freq. is 1/3 std. K-factor) 

Output freq. @ max. velocity  

( a ) 

outputs 220~240 Hz              

  ( b )

 output 73~80 Hz  

Transmission distance 

1000 metres ( 3300 feet ) maximum 

Wiring (standard) 

5 core, screened cable  

length 1.5m/5ft (DP490) or 1m/3ft (DP525) 

Protection class 

IP68  submersible ( Nema 6X ) 

Conduit entry (terminal box) 

M20 x 1.5 

Shipping Weight 

1.2 kg ( 2.7 lbs.)                 1.5 kg ( 3.3 lbs.) 

Summary of Contents for 490

Page 1: ...INSTRUCTION MANUAL BI DIRECTIONAL INSERTION METER LOCK ...

Page 2: ...ffect 7 3 2 Optional Hall Effect Reed switch output 8 3 3 Optional Reed switch output 8 3 4 Instrument cable installation requirements 9 3 5 Pulse output selection standard outputs 9 3 6 QP Quadrature pulse output option 10 3 7 Bi directional flow using QP option 10 3 8 Connection to family instruments 11 4 0 CALIBRATION K factor for meter 4 1 K factors for common pipe ID sizes 575mm 12 4 2 K fact...

Page 3: ... details 2 After screwing the Insertion Meter in place ensure the flow alignment mark located on the top positioning collar of the meter aligns with the flow in the pipe refer page 5 This ensures the paddle is correctly aligned to the flow Note The meter is bi directional so a flow direction arrow is not provided 3 Calculate and adjust the height of the Insertion Meter refer page 5 4 Electrical In...

Page 4: ...e rotor pass by pulse sensors within the transducer body inturn this produces frequency outputs proportional to flow rate 1 4 Specifications subject to change without notice Model 490 525 Suits pipe sizes 40mm 900mm 50mm 2500mm 1 5 36 inches 2 100 inches Flow range 0 25 6300 litres sec 0 4 49000 litres sec 4 99600 USGPM 6 780000 USGPM Process connections 1 5 or 2 NPT or BSPT 2 NPT or BSPT Velocity...

Page 5: ...bstructions such as pumps valves or strainers will require longer straight runs before and after the Insertion Meter 2 2 Meter installation orientation Cut a 40mm diameter hole 1 6 on either the 2 10 or 12 o clock positions of the pipe If there is any likelihood of air entrainment in a horizontal pipe do not locate the flow transducer in the 12 o clock position Install a female threaded weld on fi...

Page 6: ...1 as required so that the distance between the top of the hex adaptor 2 and the top of the positioning collar 3 equals your calculated distance A for DP490 or AA for model DP525 Retighten the height adjustment nuts 1 2 4 Flow direction orientation The unit is bi directional however the paddle must be aligned with the direction of flow Using a 2mm hex key Allen key unlock the locking screw located ...

Page 7: ...ll valve 2 5 Hot tap installation model DP525 IMPORTANT Before removing the DP525 from the installation first withdraw the Meter body to the maximum distance allowed by the three height adjusting threaded rods This will enable the isolating valve to be fully closed without damaging the paddle FLOW 2 Nipple 2 Weld O let threadolet 2 ball or gate isolation valve Allow min 40mm I D to clear metering ...

Page 8: ...etic sensor and high temperature output options A1 A2 B2 B1 b1 VOLTAGE PULSE OUTPUT Screen Yellow Green SQUARE WAVE PULSE OUTPUT Red VDC supply White Sig output Black 0v ground Terminal box option terminal connections Pull up resistor required they are generally incorporated in most receiving instruments Hex adaptor Rotor Body Positioning collar Height adjustment nuts 1 5 or 2 BSPT or NPT LOCK ...

Page 9: ...approved I S barrier Note The Reed switch produces 1 3rd the normal pulse output value eg 1 3 the standard K factor Conductor color coding also applies to the Non magnetic sensor and high temperature output options A1 A2 B2 B1 b1 REED SWITCH OUTPUT Screen Yellow Green SQUARE WAVE OUTPUT Red VDC supply White Sig output Black 0v ground Terminal box option terminal connections Pull up resistor requir...

Page 10: ...tres 3300 ft An optional intrinsically safe I S Reed Switch only output is available see page 7 Voltage pulse pulse wire output A self generating pulse output which produces a strong 1 5 volt voltage spike of approximately 10 micro second duration with no dependence on rotor speed Square Wave Pulse Hall sensor also applies to non magnetic QP Hall outputs An NPN open collector transistor pulse outp...

Page 11: ...faced with the Pulse Discriminator Module PD2 The PD2 accepts the Quadrature pulse inputs from these will discriminate between forward reverse flow Two individual proportional pulse outputs can then be sent to appropriate totalising registers or an add and subtract flow rate totaliser It is important to note that the Quadrature Pulse option has the same pulse resolution pulses unit volume as a sta...

Page 12: ...tion Meter cable should not be run with other high energy cables clause 3 3 yellow Screen Screen yellow green 6 5 4 3 2 1 7 0V ground 8 9 10 11 12 13 14 DIP switches all in the OFF position 3 2 1 O N RATE TOTALISER ECOBATCH INSTRUMENTS 1 2 3 DIP switches 2 3 to be in the OFF position ON green Vdc gnd 4 5 6 BT11 BATTERY TOTALISER gnd 1 2 3 RATE TOTALISER RESET PROGRAM ENTER gal RUN ACCUM TOTAL STOP...

Page 13: ...uare wave pulse output Pulses per litre 28647 pipe ID mm Pulses per M 28647000 pipe ID mm Pulses per US gallon 168 14 pipe ID inches Pulses per Imp gallon 201 94 pipe ID inches Pipe detail NB pipe ID inches mm p litre p m3 p USgal p litre p m3 p USgal 1 5 40 9 18 678 18678 70 695 21 524 21524 81 468 2 52 6 11 238 11238 42 534 12 818 12818 48 517 2 5 62 7 7 880 7880 29 824 8 899 8899 33 682 3 78 0 ...

Page 14: ...h 100mm ID A 24 0 2 pulses litre 1273 2 x 24 0 10000 3 056 p litre K factor for m multiply by 1000 eg K 3056 p m K factor for megalitres multiply by 1000000 eg K 3056000 p megalitre NOTE K factors for Reed Switch output option are 1 3 the standard factors of voltage pulse output 22 22 5 23 23 5 24 24 5 25 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460 A pipe I...

Page 15: ... gal 24 51 x A from graph pipe ID inches Example b K factor for 10 pipe 1 from graph 10 ID A 7 01 2 pulses gal 24 51 x 7 01 100 1 718 p gal NOTE K factors for Reed Switch output option are 1 3 the factors of the square wave pulse output 6 8 6 9 7 0 7 1 7 2 7 3 7 4 7 5 7 6 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 A pipe ID inches Pipe ID 19 5 above A 6 86 ...

Page 16: ... a full insight into flow conditioning for inferential flow devices The general rule is to have a minimum of 10 diameters of straight pipe runs before the flowmeter upstream and 5 diameters after downstream These straight runs must not contain any other items such as Valves Bends Tees Probes or Reducers etc Any Valves immediately outside the straight runs require doubling of the straight pipe run ...

Page 17: ...Installation Guidelines 16 3 PRECAUTIONS FOR AVOIDING NON UNIFORM FLOW 4 RECOMMENDED LENGHTS OF STRAIGHT PIPE FOR TYPICAL CASES 5 RECOMMENDED SOLUTIONS FOR RESTRICTED CASES ...

Page 18: ...NOTES ...

Page 19: ...NOTES ...

Page 20: ...IM DP 1716 2017 Great Plains Industries Inc All Rights Reserved Great Plains Industries Inc 888 996 3837 FLOMEC net ...

Reviews: