![FLIR EXTECH SDL310 User Manual Download Page 11](http://html1.mh-extra.com/html/flir/extech-sdl310/extech-sdl310_user-manual_2306027011.webp)
11
SDL310-EU-EN-V1.1-6/11
Electrical Specifications
(Ambient Temperature 23°C ± 5°C)
Air velocity
Measurement Range
Resolution Accuracy
m/S
0.4 - 25.0 m/s
0.1 m/S
± (2% + 0.2 m/S)
Km/h
1.4 - 90.0 km/h
0.1 Km/h
± (2% + 0.8 Km/h)
mph (Mile/h)
0.9 - 55.9 mile/h
0.1 mph
± (2% + 0.4 mph)
Knots
0.8 - 48.6 knots
0.1 Knots
± (2% + 0.4 Knots)
FPM (Ft/min)
80 - 4930 ft/min
1 FPM
± (2% + 40 FPM)
Air temperature
Measuring Range
0 to 50°C (32 to 122°F)
Resolution
0.1°C (0.1 °F)
Accuracy
± 0.8°C (1.5°F)
Type K/J Thermocouple Thermometer Temperature
Sensor Type Resolution Range
Accuracy
0.1°C
-50.0 to 1300.0°C
-100.0° to -50.1C
± (0.4%+ 0.5°C)
± (0.4%+ 1°C)
Type K
0.1°F
-58.0 to 2372.0°F
-148.0 to -58.1°F
± (0.4%+ 1°F)
± (0.4%+ 1.8°F)
0.1°C
-50.0 to 1200.0°C
-100.0 to -50.1°C
± (0.4%+ 0.5°C)
± (0.4%+ 1°C)
Type J
0.1°F
-58.0 to 2192.0°F
-148.0 to -58.1°F
± (0.4%+ 1°F)
± (0.4%+ 1.8°F)
Note: Above specifications tested under an environmental RF Field Strength lower than 3 V/M and a
frequency lower than 30 MHz
Copyright
©
2011 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com