21.1
General
The screening alarm can be used, for example, at airports to detect passengers with ele-
vated body temperatures, which may indicate the presence of a fever.
The screening alarm can also be used to detect temperature anomalies in a series of in-
spected objects in a similar/fixed setup.
Activating the screening mode will turn on a measurement box and screening data in the
result table.
The sampled average temperature.
The alarm temperature.
The measured temperature.
The alarm will trigger when the measurement box measures a temperature higher than the
alarm temperature. The alarm temperature is, in turn, the sum of a specified allowed devia-
tion and a sampled average value.
WARNING
Make sure that the laser is disabled when you point the camera at the face of a person. The laser beam
can cause eye irritation. Disable the laser by selecting
(
Settings
) >
Device settings
>
Lamp & laser
>
Disable all
.
21.2
Procedure
Follow this procedure:
1. Push the navigation pad to display the menu system.
2. Select
(
Settings
) and push the navigation pad. This displays the
Settings
menu.
3. Use the navigation pad to select
Recording mode
>
Screening
.
4. Push the navigation pad. This displays a dialog box where you can define the settings
for the alarm:
•
Allowed deviation
: The allowed deviation from the sampled average.
•
Alarm sound
: Applicable values are
Beep
or
No sound
.
5. Push the navigation pad. This closes the dialog box.
6. Push the back button
repeatedly to exit the
Settings
menu.
7. Aim the camera toward a point of interest. The object should be within the frame of the
measurement box.
8. Push and hold the programmable button
to reset the sampled average.
9. Push the programmable button
to sample.
10. Aim the camera toward more points of interest. Sample 10 times to build up a sample
base by pushing the programmable button
.
The alarm is now set up and ready to use. Occasionally record a few samples if the
alarm is used for a long time or if the conditions change.
#T810190; r. AJ/46209/46209; en-US
99
Summary of Contents for E75
Page 2: ......
Page 3: ...User s manual FLIR Exx series T810190 r AJ 46209 46209 en US iii ...
Page 4: ......
Page 52: ...Handling the camera 11 T810190 r AJ 46209 46209 en US 40 ...
Page 55: ...Handling the camera 11 11 15 Lanyard strap 11 15 1 General T810190 r AJ 46209 46209 en US 43 ...
Page 248: ...Mechanical drawings 28 See next page T810190 r AJ 46209 46209 en US 236 ...
Page 250: ...CE Declaration of conformity 29 See next page T810190 r AJ 46209 46209 en US 238 ...
Page 251: ......
Page 256: ...Application examples 30 T810190 r AJ 46209 46209 en US 244 ...
Page 297: ... T810190 r AJ 46209 46209 en US 285 ...
Page 299: ......