FT-13 Technical Manual, Rev. 1.40 March 2013
Page 53 of 56
A
PPENDIX
6:
E
RROR
T
ABLE
The FT-13 Weighing Controller has been designed as very reliable and virtually error free instruments.
However if an error occurs do not attempt to repair the equipment before you understand what caused the
error. Note the problems you have with your instrument and the error messages shown on the display. Then try
to solve the problem according to the error table given below.
Error
Code
Description
Actions to take / Possible cause
Under
Weight is too low
Check the load;
Load cell or instrument could be defective
Over
Over Load
Adc Out
Input signal outside the input range
Check the load;
Check the calibration;
Load cell or instrument could be defective
E StbL
Stability error
- The scale must be stable for zeroing, taring, tolerance control, etc.
- Increase the maximum settling time (parameter [506])
- Increase the zeroing delay (parameter [50F])
- Increase the range of the motion detector (parameter [206])
Err 1
ADC error
Re-energize indicator;
Contact FLINTEC
Err 2
Err 3
Indicator cannot get calibrated
Check load cell and load; then repeat the calibration
Err 10
EEPROM error
Configure the instrument; EEPROM defective
Err 20
Calibration error
Calibrate the indicator
Err 21
Configuration error
Configure the indicator
Err 27
Indicator is not calibrated
Calibrate the indicator
Err 28
Date & Time Error
Adjust Date & Time;
Replace the battery
Err 30
Processor Error
Contact FLINTEC
Err 34
Indicator cannot get calibrated
Load cell signal is negative , very low or too high
Err 35
Calibration Error
Calibration test weight is too small;
Check load cell connections
Err 37
Scale unstable
Wait until scale becomes stable;
Check ground connection
Err 51
Filling programme was changed (previous
process parameters have been cleared)
Press the <Enter> key; then setup all parameters for the new
programme
Err 61
U10 (eeprom) is not installed
Contact FLINTEC
Err 62
U8 (eeprom) is not installed
E XXXX
Hardware error