FAA-26, Technical Manual, Rev. 1.2.0, December 2020
Page
6
of
14
Features
•
Minimized zero and span drifts because of its microcontroller technology
•
24-bits ADC and 16-bits DAC converters.
•
Long time stability and low temperature drifts eliminates the frequent
readjustment period.
•
Very easy and user-friendly digital adjustment via rotary switches located on
the front of the instrument.
•
All instruments are adjusted to 0
– 10 V and to 4 – 20 mA analogue output
ranges for 0
– 10 mV load cell signal range in production as default. (only for
2mV/V load cells)
•
Calibration at PLC does not require readjustment after changing instrument
because of matching in the production.
TECHNICAL SPECIFICATIONS
Analogue input range
0 mV to 20 mV
Minimum input range
< 1 mV
Linearity
< % 0.01
Temperature drift
< 0.007 % FSR / ºC
Converters
24 bit Delta-Sigma ratiometric ADC with integral analogue and
digital filters; 16 bit very low drift DAC
Internal resolutions
16 000 000 counts ADC
External resolution
Analogue output changes up to 65000 steps
Calibration
With rotary switches in the front with any test load.
Preadjusted and matched instrument for calibration at PLC
Analogue outputs
Current output for 4-20 mA ;
Voltage output for 0-10 VDC.
Max. cable length
300 meter (analogue output)
Max. load resistance
(current output)
500 Ω
Load cell excitation
5 VDC
Number of Load Cells
Up to 4 units of 350
or 12 units of 1100
(min. 85
)
Power supply
12 to 28 VDC 0.2 A
Operation Temperature Between -10 °C and +45 °C at 85% RH max, non-condensing
EMC Immunity
Class E2
Enclosure
Polyamide, for DIN-rail mount, IP20
Dimensions
Front Width:22,5 mm, Front Length: 99 mm , Height:114,5mm,