Clave 23+ Instruction Manual v1.4 

5.1.4 Filling the distilled water tank 

Open the cover and fill the tank with distilled water. When you hear the beep signal, it 

means that the water level has reached the maximum level. The icon 

will be 

displayed. Please stop filling immediately. 
 
 

 

 

 

 

The water level should not exceed this port. 

 

 

5.2 Preparation of Sterilization Materials 

 

For the most effective sterilization and to preserve the instrument and tools, please follow 
the suggestions below: 
* Arrange the instruments of different materials on different trays or with at least 3cm of 
space between them. 
* Always insert a sterilization paper or cloth between the tray and instrument to avoid 
direct contact between the different materials. 
* Arrange the containers (glasses, cups, test-tubes, etc) on one side or in an inverted 
position to avoid water pooling. 
* Don’t stack the trays one above the other or put them in direct contact with the walls of 
the sterilization chamber. 
* Always use the instrument tray handle. 
* Wrap the instruments one by one or, if more tools have to be set in the same bag, verify 
that these are made of the same material. 
* Don’t use metallic clips, pins, or other similar things as this jeopardizes the 
maintenance of the sterilizer. 
* Don’t overload the trays over the stated limit (see Appendix 2). 

 

5.3 Selecting the Sterilization Program 

 
 

5.3.1 LCD: The panel displays the cycle temperature, 

 

pressure, error code, sterilization state and program. 

 

5.3.2 Select button  . Select an item and save an adjustment. 

 

5.3.3 UP button  . Press UP button to select a program or  

 

adjust and setup the parameter. 

 

5.3.4 DOWN button  . Press DOWN button to select a

 

program or adjust and setup the parameter.  

 

5.3.5 Start button 

. Press this button to start the sterilization 

 

cycle.   To stop a cycle, press and hold this button for 5 seconds. 

 

5.3.6 SET button. Not used. 

Summary of Contents for Clave 23+

Page 1: ...1 Clave 23 Instruction Manual v1 4 Clave 23 Instruction Manual Steam Sterilizer Distributed by...

Page 2: ...ode E88 will be displayed after 1 year of use or after a certain number of cycles whichever comes first This is programmed into the maintenance screen The Clave 23 is the same as a Clave 23 except it...

Page 3: ...ials 8 5 3 Selecting the sterilization program 8 5 4 Running the sterilization program 9 5 5 Basic set 10 5 6 Record of the cycle 10 6 Advanced settings 11 7 Maintenance 14 8 Error codes 16 9 Transpor...

Page 4: ...cedure highlight and select for any other records to be transferred to the USB stick Checking Maintenance Screen 1 Hold the Up and Down buttons while powering on Let go when you enter the interface Sc...

Page 5: ...denser exhaust 6 Used water tank drain port 14 Rating plate 7 Door handle 15 Circuit breaker 8 Door 16 Power socket Security Notice For safe operation please pay close attention to the warning symbols...

Page 6: ...imum level Operation temperature 5 C 40 C Exterior dimensions 480mm W x 445mm H x 700mm D 18 9 W x 17 5 H x 27 6 D Net weight 50kg 110 2 lbs Noise level 70db Relative humidity Maximum 80 non condensin...

Page 7: ...1 Setup 5 1 1 Open the door and remove all of the inner contents 5 1 2 Connect the power cord to an outlet of the appropriate voltage 5 1 3 Power on The switch is located on the right side of the cove...

Page 8: ...on to avoid water pooling Don t stack the trays one above the other or put them in direct contact with the walls of the sterilization chamber Always use the instrument tray handle Wrap the instruments...

Page 9: ...chamber using the tray handle 5 4 1 After the instruments are loaded you may close and lock the door by turning the door handle clockwise until it stops Caution You must turn the door handle to the m...

Page 10: ...ing time to N It will skip the drying stage 5 5 Basic Set 5 5 1 Highlight the Basic Set and press the SELECT button 5 5 2 Select the item by pressing SELECT button 5 5 3 Adjust the value by pressing t...

Page 11: ...Ambient Pressure 6 2 2 Adjust the parameter by pressing button Press the SELECT button to save Press START button to cancel and exit 6 2 3 The value is 0 2 If the machine can t enter the holding time...

Page 12: ...ht side panel The information will be automatically recorded to the USB drive after the cycle has completed The name of the file is determined by the serial number of the machine and the cycle number...

Page 13: ...50 5kPa T4 13 31 31 084 0C 009 7kPa T4 15 11 50 083 5C 009 9kPa T5 13 35 42 109 8C 102 7kPa T5 15 15 50 108 2C 101 1kPa T6 13 36 22 098 8C 009 9kPa T6 15 16 30 097 2C 009 7kPa TS 134 8C 222 5kPa TS 00...

Page 14: ...Replace the bacteria filter Every 6 12 months Replace the door seal 7 1 Clean the distilled water tank every week with isopropyl alcohol or a medical disinfectant Drain unused water 7 2 Clean the cha...

Page 15: ...ighten the sealing plate 7 4 3 Turn the nut until the sealing plate is tight If the door knob is too tight you may also turn the nut clockwise to loosen it Caution Never adjust the chamber door while...

Page 16: ...curs at 8 minutes E06 Door lock fault Make sure you have turned the door handle to the maximum position Check the door switch E09 Pressure leak takes too long to reach pressure and temperature Check i...

Page 17: ...Action Interruption of the power to the transformer Contact a service person 3 Safety valve Protection against possible sterilization over pressure Action Releases steam and restores pressure inside...

Page 18: ...mg L 1 0mg kg Silicon dioxide SiO2 1mg L 0 1mg kg Iron 0 2mg L 0 1mg kg Cadmium 0 005mg L 0 05mg kg Lead 0 05mg L 0 1mg kg Rest of heavy metals 0 1mg L 0 1mg kg Chloride 2mg L 0 1mg kg Phosphates 0 5m...

Page 19: ...apped porous material 1 25 0 40 Single wrapped porous material 1 10 0 30 Dual wrapped porous material 0 75 0 25 Single wrapped hollow material 4 00 1 25 Dual wrapped solid and hollow material 2 00 0 6...

Page 20: ...20 Clave 23 Instruction Manual v1 4...

Page 21: ...ion Manual v1 4 APPENDIX 3 WIRING DIAGRAM TP1 Steam generator temperature sensor TP2 Inner temperature sensor TP3 Chamber wall temperature sensor V1 Air release valve normally open V4 Water release va...

Page 22: ...22 Clave 23 Instruction Manual v1 4 Appendix 4 Hydraulic Diagram...

Page 23: ...must be paid for by the owner of the Autoclave Flight will not accept COD shipments If repairs are needed during the first 90 days after purchase of an autoclave and a local authorized service dealer...

Page 24: ...s regarding the document is welcomed For any comments and concerns please mail e mail or phone us at Sales and Marketing sales flightdentalsystems com 1 866 799 0517 905 799 0517 Technical Support 1 8...

Reviews: