Flight DEFENDME Aerosol Evacuator  R.1.2 

 

13 

 

if there is a blockage between 
the directional duct and the 
top cover preventing air from 
going into the main enclosure.  
After these steps, restart the 
machine and observe for one 
minute if the alarm still comes 
on. If it does than replace the 
filters.  

LED Display has “ERR” alarm, 
and the fan and motor has 
stopped working 

The fan has malfunctioned or 
the connection to the fan has 
a problem. 

Turn off the power and restart 
the machine to check whether 
the fan is working properly, if 
not check the fan and the 
connections to the fan to 
ensure they are all secure. 

9.0

 

Technical Data 

 

Manufacturer 

HR Dental Products Inc. o/a Flight Dental Systems 

Model 

DefendMe Aerosol Evacuator 

Model # 

AE-1000 

Dimensions 

16 x 9 x 18” (L x W x H) 

Duct 

Diameter 3 inch x 59 inch Length (D75 x 1500L) 

Weight 

46 lbs (21kg) 

Input 

100-120V/60hz 

Power 

250W Max 

Filter Efficiency (0.3um) 

99.97% 

Static Pressure 

3000pa 

Air Rate 

14 m/s 

Volume (Filter Included) 

3.7m

3

/ min 

Ultraviolet 

Ultraviolet lamp intensity: 2260uW/cm

3

,  

Wavelength: 265-285nm 

Electrical Safety Class 

Class I 

Applied Part 

Operating Conditions 

Use: In Enclosed Spaces 
Ambient Temperature: 5C to 40C 
Relative Humidity: <80% 
Operating Altitude <3000m above sea level 

Transport and Storage 
Conditions 

Ambient Temperature: -20C to +55C 
Relative Humidity: 20% to 80% 
Atmospheric Pressure: 70kPa to 106kPa 

Summary of Contents for AE-1000

Page 1: ...Flight DEFENDME Aerosol Evacuator R 1 2 1 DEFENDME AEROSOL EVAUCATOR USER MANUAL...

Page 2: ...ing DefendMe Aerosol Evacuator 7 5 0 User Interface 8 6 0 Operating Instructions 9 6 1 Air Volume Setting 9 6 2 Connect the Speed Controller for Air Volume Adjustment 9 6 3 Power Off 9 7 0 Maintenance...

Page 3: ...Flight DEFENDME Aerosol Evacuator R 1 2 3 1 0 Scope of DefendMe Aerosol Evacuator 1 Main Enclosure 2 Directional Duct 3 Suction Nozzle 4 Speed Controller 5 Power Cord...

Page 4: ...nt NOTE Additional information Explanation of operation and performance Serial number Catalogue number Date of Manufacturing LOT Lot Number Protective earth ground Type B applied part Fuse WEEE direct...

Page 5: ...bacteria and viruses in aerosols created during dental procedures The device will vacuum aerosols into the system and filter and disinfect any viruses or bacteria Please note that the device cannot be...

Page 6: ...y undoing the 4 latches marked by the blue arrows and remove the top cover 2 Remove the main filter assembly Remove the rubber gasket seal and filter elements in sequence Then take out the main filter...

Page 7: ...tional duct by screwing the nut onto the duct 2 Installation of the Directional Duct Plug the direction duct to the top of the cover Directional duct should be installed vertically and perfectly secur...

Page 8: ...hown in the picture 5 0 User Interface 1 Screen Display 2 Speed Increase 3 Speed Decrease 4 Power Switch Turn power On Press Power Switch to to turn on Speed Increase Press speed up key to change air...

Page 9: ...recommended to work under the condition of larger air volume 6 3 Power Off Turn off the device by pressing the volume controller power switch to OFF and the window will Do not block the air inlet to a...

Page 10: ...equency of use Comply with your local regulations for cleaning and disinfection The device should no longer be used if there are signs of damage to the device or any its Condensation may appear on the...

Page 11: ...er For cleaning of the components please use distilled water only Do not use bleach or chloride disinfectant materials Components to Disinfect Suction Nozzle Directional Duct Main Enclosure Power Cord...

Page 12: ...bly and main filter in that order Replace each filter accordingly When replacing the main filter ensure that the filter is placed correctly and that the plug for the fan is connected to the bottom of...

Page 13: ...sure they are all secure 9 0 Technical Data Manufacturer HR Dental Products Inc o a Flight Dental Systems Model DefendMe Aerosol Evacuator Model AE 1000 Dimensions 16 x 9 x 18 L x W x H Duct Diameter...

Page 14: ...use in the electromagnetic environment specified below The customer or the user of the DEFENDME should assure that it is used in such an environment Emissions test Compliance Electromagnetic environme...

Page 15: ...kV 15 kV air 8 kV contact 2 kV 4 kV 8 kV 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical...

Page 16: ...vironment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted dis turbances induced by RF fields IEC 61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM bands be tw...

Page 17: ...nity test level in the below table and meet the related requirements of IEC 60601 1 2 2014 The customer and or user should help keep a minimum distance between RF wireless communications equipment and...

Reviews: