Flight DEFENDME Aerosol Evacuator R.1.2
11
components. Please contact your authorized dealer for repair or replacement of any
components.
7.1.2 General Recommendations
•
The user is responsible for disinfecting the product before use.
•
For you own safety, please wear the appropriate personal protective equipment (PPE)
(gloves, safety glasses, gowns etc) when using and cleaning this device.
•
Use only the correct disinfection solutions or wipes as approved by your local regulators
(FDA, Health Canada) and in accordance with the IFU of the manufacturer
•
For cleaning of the components please use distilled water only.
•
Do not use bleach or chloride disinfectant materials.
Components to Disinfect
Suction Nozzle
Directional Duct
Main Enclosure
Power Cord
Speed Controller
Wipe all the surfaces with a cloth or wipe with at least 70-80% ethanol or an approved
disinfectant. Wait at least 2 minutes and repeat 5 times. Or follow the recommendations of the
disinfectant manufacturer.
Do not spray excessive ethanol or disinfection solution as it may go into the system and
damage the components inside.
7.2 Filter Replacement
7.2.1 Frequency of Filter Replacement
Components
Frequency
Durability
Flow Value (Usage)
Primary Microfiber Filter
1-3 months
Display 010 or a multiple of it
Eg. 010/020/030.